IDENTIFIKASI ARUS BERKAS ELEKTRON PADA PRA KOMISIONING MESIN BERKAS ELEKTRON (MBE) LATEKS

dokumen-dokumen yang mirip
UJI FUNGSI SISTEM PEMAYAR MESIN BERKAS ELEKTRON 300 KEV/20 MA

OPERASI MESIN BERKAS ELEKTRON (MBE) PTAPB BATAN TIPE BA 350 kev / 10 ma

PENENTUAN DOSIS RADIASI MENGGUNAKAN DOSIMETER FRICKE

PENGUJIAN SISTEM VAKUM MBE 350keV/10 ma PASCA PENGGANTIAN POMPA TURBOMOLEKUL

PENGUKURAN DOSIS RADIASI IRADIATOR GAMMA DAN MESIN BERKAS ELEKTRON DENGAN DOSIMETER CERI- CERO

Prodi Fisika FMIPA, Universitas Sebelas Maret, Surakarta.

PENENTUAN KEDALAMAN PENETRASI BERKAS ELEKTRON 800 kev DALAM GAS BUANG PLTU PADA SISTEM PENGOLAHAN GAS BUANG MENGGUNAKAN MESIN BERKAS ELEKTRON

OPTIMASI UJI KINERJA OPERASI PROTOTIP LITBANG MBE-PSTA PADA 200 KEV

BAHAN DAN METODE PENELITIAN

KAJIAN PENENTUAN KEDALAMAN PENETRASI BERKAS ELEKTRON 350 kev PADA HIDROGEL UNTUK PEMBALUT LUKA

PENGUKURAN DISTRIBUSI MEDAN MAGNET SISTEM OPTIK MBE PADA TAHAP PRA-KONSTRUKSI

SIMULASI SISTEM INTERLOCK PENGAMAN OPERASI MESIN BERKAS ELEKTRON (MBE) DENGAN PERANGKAT LUNAK BASCOM 8051

OPTIMASI KINERJA OPERASI SISTEM PEMAYAR MBE LATEKS 300 kev/20 ma

RANCANG BANGUN TRANSFORMATOR 7,2 V/200 A SEBAGAI CATU DAYA FILAMEN TABUNG TRIODA ITK 15-2 PADA GENERATOR COCKCROFT WALTON MBE LATEKS 300keV/20 ma

Pengukuran Dosis Berkas Elektron Dari Mesin Berkas Elektron Pada Biji Tumbuhan Menggunakan Spectrophotometer Genesys 5. Skripsi

PEMBUATAN RODA GIGI REDUKSI PEMUTAR VARIAK SISTEM TEGANGAN TINGGI MBE INDUSTRI LATEK

RANCANGAN AWAL PERISAI RADIASI MESIN BERKAS ELEKTRON DUET

RANCANGAN SISTEM CATU DAYA DC 2 kv/2 A UNTUK KATODA SUMBER ION SIKLOTRON 13 MeV BERBASIS TRANSFORMATOR

RANCANGAN TRANSFORMATOR 625 VA TERISOLASI PADA TEGANGAN TINGGI 300 KV UNTUK CATU DAYA FILAMEN SUMBER ELEKTRON MBE LATEKS

ANALISIS SIMULASI LINTASAN BERKAS ELEKTRON PADA IRADIATOR ELEKTRON PULSA (IEP) DENGAN VARASI GEOMETRI ELEKTRODA PEMFOKUS MENGGUNAKAN SOFTWARE

PENENTUAN DOSIS SERAP BERKAS ELEKTRONUNTUK PENGOLAHAN GAS BUANGMENGGUNAKAN MESIN BERKAS ELEKTRON (MBE)

Volume 17, November 2015 ISSN

DESAIN DAN KONSTRUKSI CORONG PEMAYAR MESIN BERKAS ELEKTRON

PROSIDING SEMINAR PENELITIAN DAN PENGELOLAAN PERANGKAT NUKLIR. Pusat Teknologi Akselerator dan Proses Bahan. Yogyakarta, 28 Agustus 2008

HUBUNGAN TEGANGAN DAN CITRA RADIOGRAFI REAL TIME PADA PESAWAT SINAR-X RIGAKU RADIOFLEX-250EGS3

RANCANG BANGUN SISTEM INTERLOK UNTUK OPERASI MBE LATEKS MENGGUNAKAN MIKROKONTROLER

RANCANGBANGUN PERANGKAT SISTEM INDIKATOR LAMPU LED PANEL KENDALI MBE 300 kev/20 ma

Analisis Persamaan Respon Dosis Thermoluminescent Dosimeter (TLD) Pada Spektrum Sinar-X Menggunakan Metode Monte Carlo

KAJIAN PENGARUH WARNA DAN JARAK LAMPU PENGAMAN TERHADAP HASIL RADIOGRAF

PEMBUATAN FARADAY CUP BERBASIS LAYAR PENDAR SEBAGAI SENSOR BERKAS PARTIKEL ELEKTRON

PENGUJIAN TANGKI BERTEKANAN UNTUK SISTEM PEMERCEP AT PADA MBE LATEKS

oleh Werdi Putra Daeng Beta, SKM, M.Si

PENGUKURAN DOSIS RADIASI RUANGAN RADIOLOGI II RUMAH SAKIT GIGI DAN MULUT (RSGM) BAITURRAHMAH PADANG MENGGUNAKAN SURVEYMETER UNFORS-XI

PROSIDING SEMINAR PENELITIAN DAN PENGELOLAAN PERANGKAT NUKLIR Pusat Teknologi Akselerator dan Proses Bahan Yogyakarta, 28 Agustus 2008

EKSPERIMEN UJI PADA DAYA TINGGI DARI HEAD SUMBER ION UNTUK SIKLOTRON

UJI VAKUM BEJANA NITRIDASI PLASMA

SIMULASI PENGARUH DAYA TERDISIPASI TERHADAP SISTEM PENDINGIN PADA BEJANA TEKAN MBE LATEKS

PERHITUNGAN ORBIT AWAL BERKAS PROTON PADA CENTRAL REGION SIKLOTRON

KAJIAN PENGARUH WARNA DAN JARAK LAMPU PENGAMAN TERHADAP HASIL RADIOGRAF

PENGUKURAN DOSIS PAPARAN RADIASI DI AREA RUANG CT SCAN DAN FLUOROSKOPI RSUD DR. SAIFUL ANWAR MALANG. Novita Rosyida

RANCANGAN SISTEM PENGATUR CATU DAYA SUMBER ELEKTRON TIGA ELEKTRODA MBE LATEKS BERBASIS PLC

VERIFIKASI PENENTUAN LAJU DOSIS SERAP DI AIR BERKAS FOTON 6 MV DAN 10 MV PESAWAT PEMERCEPAT LINIER MEDIK CLINAC 2100 C MILIK RUMAH SAKIT

PENGUJIAN LINIERITAS KELUARAN PEMBANGKIT ARUS SINAR X MENGGUNAKAN STEPWEDGE SKRIPSI. Evi Yusita Nim

PENGARUH POLARITAS MEDAN LISTRIK EKSTERNAL DAN SUDUT POLARISASI LASER DIODA UNTUK PENGAMATAN EFEK KERR

DAFTAR KELENGKAPAN DOKUMEN YANG HARUS DILAMPIRKAN

Youngster Physics Journal ISSN : Vol. 2, No. 1, April 2013, Hal 27-34

ANALISIS KUALITAS RADIASI DAN KALIBRASI LUARAN BERKAS FOTON 6 DAN 10 MV PESAWAT PEMERCEPAT LINIER MEDIK VARIAN CLINAC CX 4566 ABSTRAK

STANDAR NASIONAL INDONESIA (SNI) BIDANG NUKLIR

SIMULASI DAN MONITORING SISTEM INTERLOCK MESIN BERKAS ELEKTRON PTAPB BATAN DENGAN PERANGKAT SUARA

HUBUNGAN ANTARA LAJU DOSIS SERAP AIR DENGAN LAPANGAN RADIASI BERKAS ELEKTRON PESAWAT PEMERCEPAT LINIER MEDIK ELEKTA

OPTIMALISASI DOSIS RADIASI SINAR-X TERHADAP PROYEKSI PA (POSTERO-ANTERIOR) DAN LAT (LATERAL) PADA TEKNIK PEMERIKSAAN FOTO THORAX SKRIPSI

Uji Kesesuaian Pesawat Fluoroskopi Intervensional merek Philips Allura FC menggunakan Detektor Unfors Raysafe X2 di Rumah Sakit Universitas Andalas

PENGARUH VARIASI AIR GAP TERHADAP DOSIS SERAP PENYINARAN BERKAS ELEKTRON PADA PESAWAT LINAC SIEMENS / PRIMUS M CLASS 5633

UJI KESESUAIAN PESAWAT CT-SCAN MEREK PHILIPS BRILIANCE 6 DENGAN PERATURAN KEPALA BAPETEN NOMOR 9 TAHUN 2011

PENENTUAN KAPASITAS PENGOLAHAN GAS BUANG BATU BARA DENGAN MENGGUNAKAN MESIN BERKAS ELEKTRON

BAB II LANDASAN TEORI

PENENTUAN DAYA SERAP APRON DARI KOMPOSIT KARET ALAM TIMBAL OKSIDA TERHADAP RADIASI SINAR- X

PEMBUATAN TABUNG DETEKTOR GEIGER MULLER TIPE JENDELA SAMPING

Analisis Pengaruh Sudut Penyinaran terhadap Dosis Permukaan Fantom Berkas Radiasi Gamma Co-60 pada Pesawat Radioterapi

PENGUKURAN BENTUK PROFIL BERKAS ELEKTRON DARI SUMBER ELEKTRON TIPE PIERCE MENGGUNAKAN SENSOR TABUNG TV BEKAS

PENENTUAN KEMBALI KOMPOSISI KOMPOSIT KARET ALAM TIMBAL OKSIDA SEBAGAI PERISAI RADIASI SINAR-X SESUAI KETENTUAN BAPETEN

Penentuan Dosis Gamma Pada Fasilitas Iradiasi Reaktor Kartini Setelah Shut Down

PENGUJIAN AWAL SISTEM INSTRUMENTASI & KENDALI UNIT VAKUM MESIN BERKAS ELEKTRON 350 kev/10 ma TIPE REMOTE MANUAL

EVALUASI DOSIS RADIASI EKSTERNAL PEKERJA PUSAT RADIOISOTOP DAN RADIOFARMAKA ( PRR )

SUMBER ELEKTRON TIGA ELEKTRODA MENGGUNAKAN TABUNG PEMERCEPAT NEC UNTUK MBE-LATEKS 300KV/20 MA

KEPALA BADAN PENGAWAS TENAGA NUKLIR REPUBLIK INDONESIA

PEREKAYASAAN PERISAI RADIASI TIROID MENGGUNAKAN KOMPOSIT LATEKS CAIR TIMBAL OKSIDA DENGAN TEKNOLOGI ULTRA SONIK DAN SUHU SUPER KRITIS

KONSTRUKSI DAN UJI FUNGSI SISTEM INSTRUMENTASI DAN KENDALI PERANGKAT VAKUM MBE 300 kev/20 ma

BAB IV HASIL PENGAMATAN

KARAKTERISASI DIFRAKSI SINAR X DAN APLIKASINYA PADA DEFECT KRISTAL OLEH: MARIA OKTAFIANI JURUSAN FISIKA

ANALISIS KUALITAS RADIOGRAFI PADA OBJEK BERGERAK DAN OBJEK TIDAK BERGERAK DENGAN MENGGUNAKAN VARIASI EKSPOSE SKRIPSI

PROSIDING SEMINAR PENELITIAN DAN PENGELOLAAN PERANGKAT NUKLIR. Pusat Teknologi Akselerator dan Proses Bahan. Yogyakarta, 28 Agustus 2008

PENGARUH TEGANGAN TABUNG (KV) TERHADAP KUALITAS CITRA RADIOGRAFI PESAWAT SINAR-X DIGITAL RADIOGRAPHY (DR) PADA PHANTOM ABDOMEN

DESAIN SISTEM PENDINGIN TRANSFORMATOR FREKUENSI TINGGI PADA MESIN BERKAS ELEKTRON 300 kev/20 ma

1BAB I PENDAHULUAN. sekaligus merupakan pembunuh nomor 2 setelah penyakit kardiovaskular. World

KARAKTERISASI TiO 2 (CuO) YANG DIBUAT DENGAN METODA KEADAAN PADAT (SOLID STATE REACTION) SEBAGAI SENSOR CO 2

APLIKASI IRADIASI MESIN BERKAS ELEKTRON UNTUK DISINFESTASI SERANGGA Tribolium castaneum (Herbst) PADA TEPUNG TERIGU RINDY PANCA TANHINDARTO

Analisa Kualitas Sinar-X Pada Variasi Ketebalan Filter Aluminium Terhadap Dosis Efektif

EVALUASI DOSIS RADIASI EKSTERNAL PEKERJA PUSAT RADIOISOTOP DAN RADIOFARMAKA ( PRR )

FORMULIR PERMOHONAN IZIN PEMANFAATAN SUMBER RADIASI PENGION UNTUK KEGIATAN PENGGUNAAN DALAM RADIOGRAFI INDUSTRI

EVALUASI DOSIS RADIASI INTERNAL PEKERJA RADIASI PT-BATAN TEKNOLOGI DENGAN METODE IN-VITRO

PENGUKURAN DOSIS RADIASI PADA PASIEN PEMERIKSAAN PANORAMIK. Abdul Rahayuddin H INTISARI

UJI FUNGSI ALAT PENGENDALI SUHU TIPE TZ4ST-R4C SEBAGAI PERANGKAT PENGKONDISIAN SINYAL

SPEKTROFOTOMETRI. Adelya Desi Kurniawati, STP., MP., M.Sc.

UJICOBA SISTEM ELEKTRODE SUMBER ELEKTRON BERBASIS KATODE PLASMA

EVALUASI TEBAL DINDING RUANGAN PESAWAT LINEAR ACCELERATOR (LINAC) SINAR-X DI INSTALASI RADIOTERAPI RUMAH SAKIT UNIVERSITAS HASANUDDIN

SIMULATION FOR RADIATION SHIELDING DESIGN OF EBM-LATEX USING MCNP5

RANCANGBANGUN PERANGKAT LUNAK KERANGKA PROGRAM PANEL KENDALI MBE 300 kev/20 ma BERBASIS KOMPUTER

PENENTUAN NILAI KOEFISIEN SERAPAN BAHAN DAN DOSIS RADIASI PADA VARIASI KOMBINASI KAYU DAN ALUMINIUM

PEMERIKSAAN KUALITAS BOOM FOOT MENGGUNAKAN TEKNIK UJI TAK RUSAK

EVALUASI KESELAMATAN RADIASI PENGUNJUNG DI TEMPAT PENYIMPANAN SEMENTARA LIMBAH RADIOAKTIF

PENENTUAN PARAMETER KOMPONEN PERANGKAT UJI SUMBER ION SIKLOTRON

PERANCANGAN RUANGAN RADIOGRAFI MEDIK DI SEKOLAH TINGGI TEKNIK NUKLIR

MAKALAH FABRIKASI DAN KARAKTERISASI XRD (X-RAY DIFRACTOMETER)

Gravitasi Vol. 15 No. 1 ISSN:

PENENTUAN SPHERICITY DAN DISTRIBUSI INTENSITAS BERKAS ELEKTRON DARI SUMBER ELEKTRON TIPE PIERCE BERBASIS MATLAB

ANALISIS PENGUKURAN LINIERITAS KELUARAN PADA PESAWAT SINAR-X RADIOGRAFI UMUM DI RSUD LANGSA. Hadi SAPUTRA NIM :

FORMULIR PERMOHONAN IZIN PEMANFAATAN SUMBER RADIASI PENGION UNTUK KEGIATAN : IMPOR ZAT RADIOAKTIF UNTUK KEPERLUAN SELAIN MEDIK

SIMULASI EFISIENSI DETEKTOR GERMANIUM DI LABORATORIUM AAN PTNBR DENGAN METODE MONTE CARLO MCNP5

PENGARUH SUDUT GANTRI TERHADAP KONSTANSI DOSIS SERAP DI AIR PESAWAT TELETERAPI Co-60 XINHUA MILIK RUMAH SAKIT dr. SARJITO YOGYAKARTA

Transkripsi:

IDENTIFIKASI ARUS BERKAS ELEKTRON PADA PRA KOMISIONING MESIN BERKAS ELEKTRON (MBE) LATEKS Sukaryono, Rany Saptaaji, Suhartono, Heri Sudarmanto -BATAN, Yogyakarta Email : ptapb@batan.go.id ABSTRAK IDENTIFIKASI ARUS BERKAS ELEKTRON PADA PRA KOMISIONING MESIN BERKAS ELEKTRON (MBE) LATEKS. Telah dilakukan identifikasi arus berkas elektron pada pra komisioning MBE Lateks. Identifikasi arus berkas elektron dilakukan untuk mengetahui distribusi arus berkas elektron yang mengenai target. Identifikasi dilakukan dengan mengoperasikan MBE Lateks pada kondisi tegangan pemercepat 208 kv, arus berkas 25 µa. Untuk mengidentifikasi bahwa berkas elektron sudah dihasilkan, maka pada sistem target dipasang dosimeter go no go dan CTA film. Dosimeter gonogo dipasang untuk mengetahui secara visual apakah sudah dihasilkan arus berkas elektron atau belum, apabila terjadi perubahan warna dari kuning menjadi merah maka menandakan bahwa sudah dihasilkan arus berkas elektron. CTA film digunakan untuk mendeteksi besarnya dosis arus berkas elektron yang mengenai target. Pemantauan distribusi dosis dilakukan pada tiga titik pantau yaitu selatan, tengah dan utara dengan luasan target (60 x 6) cm. Dosis arus berkas elektron tertinggi berada pada tengah luasan target yaitu 14,8 kgy,dosis pada sisi selatan dan utara masing-masing 11,8 kgy dan 8,5 kgy.dari data pengukuran keseragaman dosis sepanjang window MBE, diperoleh D maks /D min = 1,74, dengan demikian keseragaman dosis sepanjang window MBE tidak homogen. ABSTRACT IDENTIFICATION OF ELECTRON BEAM CURRENT IN THE PRE COMISSIONING OF LATEX ELECTRON BEAM MACHINE (EBM). Identification of electron beam current in the pre comissioning of latex EBM has been done. Identification of the electron beam current is conducted to determine the electron beam current distribution on the target. Identification is done by operating the latex elctron beam machine on 208 kv accelerating voltage and 25 µa beam current. To identify that the electron beam is generated, go no go dosimeter and CTA film are placed in the target system. Go no go dosimeter is placed to determine wheather the electron beam current is generated by visual, if the color turn from yellow into red then the electron beam current is generated. CTA film is used to detect the magnitude of the dose of electron beam current that hit the target. The monitoring of dose distribution is carried out in three monitoring spot that is south, middle and north of the target with the target area (60 x 6) cm. The highest electron beam current dose is at the middle of the target area that is 14.8 kgy, the dose in the south and north side are 11.8 kgy and 8.5 kgy respectively. From the measurement of dose uniformity along electron beam machine window, it was obtained D maks /D min = 1.74 thus the dose uniformity throughout the window was not homogeneous. PENDAHULUAN M ulai tahun 2005 Pusat Teknologi Akselerator dan Proses Bahan mempunyai program rancangbangunmesin berkas elektron (MBE) Lateks dengan kapasitas 300 kev/20 ma, program ini diawali dengan pembuatan rancangan dasar dilanjutkan dengan pembuatan rancangan detil serta realisasi pembuatan (1). Pada tahun 2012 program Sukaryono, dkk. ISSN 1410 8178 Buku II hal. 443

rancangbangun MBE Lateks memasuki tahap pengujian, keberhasilan dari program rancang bangun dapat terukur apabila elektron yang dibangkitkan oleh sumber elektron bisa dipercepat pada ruang hampa dan diatur mengenai target (2). Berkenaan dengan hal tersebut maka dilakukan pengukuran dosis serap untuk menentukan distribusi dosis berkas elektron sepanjang window MBE. Untuk mengetahui berkas elektron sudah sampai target atau belum dapat dilakukan dengancara kualitatif maupun kuantitatif. Yang dimaksud cara kualitatif adalah memastikansudah terjadi proses radiasi atau tidak pada target, yaitu dengan menggunakan indikator radiasi yang biasa disebut dosimeter gonogo. Dosimeter ini berupa zat warna yang peka terhadap radiasi dan apabila terkena radiasi akan berubah warna menjadi lebih gelap sebanding dengan dosis radiasi yang mengenainya. Sedangkan cara kuantitatif bertujuan untuk mengetahui besaran dosis serap berkas elektron yang terkena pada target. Dalam kegiatan ini pengukuran distribusi dosis dilakukan pada tiga posisi yaitu selatan, tengah dan utara luasan window MBE. Distribusi dosis serap sepanjang window MBE dapat digunakan untuk menetapkan dosis maksimum dan dosis minimum sebagai parameter keseragaman dosis (dose uniformity). Hal ini perlu dilakukan karena keseragaman dosis merupakan parameter pengendalian dosis dalam melakukan iradiasi suatu bahan agar diperoleh kualitas bahan yang seragam/homogen. Dalam pengukuran dosis yang terserap pada bahan diperlukan alat dosimeter. Dosimeter film, seperti Celulosa Triacetat (CTA) sangat tepat digunakan sebagai kendali mutu proses iradiasi berkas elektron (3). Evaluasi dosis serap dilakukan berdasarkan perubahan rapat optik (optical density) atau absorban pada CTA sebelum dan sesudah diradiasi. Perubahan rapat optik/absorban diukur dengan SpectrophotometerGenesys 5 pada panjang gelombang 280 nm. Gambar 1. Skema Mesin Berkas Elektron Lateks (300 kev/20 ma) TATA KERJA Identifikasi berkas elektron dilakukan dengan mengoperasikan MBE Lateks pada tegangan pemercepat tertentu sehingga dihasilkan arus berkas elektron pada target. Bahan dan Peralatan Dosimeter Celulosa Triacetat (CTA) dan dosimeter go no go. Spectrophotometer Genesys 5 buatan Amerika Mesin Berkas Elektron 300 kev/20 ma (MBE) Lateks Selotip, spidol permanen, tisu, sarung tangan, pinset, gunting Pakaian kerja, surveymeter, film badge. Identifikasi Arus Berkas Elektron Seluas Jendela Pemayar Menyiapkan papan target sesuai dengan ukuran lebar window dan ditempel dosimeter go no go serta CTA pada tiga posisi target (utara, tengah dan selatan). Menempatkan papan target yang sudah dipasang dosimeter go no go dan CTA di bawah window MBE dengan jarak 2,5 cm. MBE dioperasikan pada tegangan pemercepat dan arus berkas elektron masing-masing 208 kv dan 25 µa untuk diiradiasikan ketarget. Mengamati apakah ada perubahan warna pada dosimeter go no go dan dilanjutkan melakukan persiapan pembacaan rapat optik/absorban dosimeter CTA. Mengkondisikan dosimeter CTA yang telah diiradiasi dengan MBE Lateks selama kurang lebih 2 jam. Mengukur/membaca rapat optik/absorban dosimeter CTA setelah diiradiasi menggunakan Spectrophotometer Genesys 5. Perhitungan dosis serap menggunakan kurva kalibrasi dosis vs respon dosimeter CTA. HASIL DAN PEMBAHASAN Dalam identifikasi berkas elektron dilakukan dengan mengoperasikan MBE Lateks pada tegangan pemercepat 208 kv sehingga dihasilkan arus berkas elektron 25 µa pada kondisi vakum 5.10-6 mbar. Berkas elektron diiradiasaikan ke target yang sudah dipasang dosimeter go no go dan dosimeter CTA. Dari hasil iradiasi terlihat pada dosimeter go no go ada perubahan warna dari kuning menjadi merah kecoklatan, hal ini menandakan ada berkas elektron yang sampai ke target. Sedangkan untuk mengukur besarnya dosis di target menggunakan dosimeter CTA. Dosimeter CTA yang digunakan adalah tipe FUJI FTR 125 buatan Jepang berupa pita panjang warna bening dengan dimensi lebar 8 mm, tebal 0,125 mm, Buku II hal. 444 ISSN 1410 8178 Sukaryono, dkk

densitas 1,298 gr/cm 3 dan dapat dibaca pada panjang gelombang 280 nm. Sebelum melakukan pengukuran dosis serap menggunakan dosimeter CTA tipe FUJI FTR 125 ada beberapa hal yang harus diperhatikan agar tidak merubah kondisi dosimeter CTA, yaitu pada kedua permukaan tidak boleh kotor, terkena minyak/lemak, debu, tidak boleh dipegang langsung tangan dan tidak boleh terlipat, karena kondisi tersebut akan mempengaruhi data hasil pengukuran rapat optik/absorban. Alat yang digunakan untuk membaca nilai rapat optik/absorban sebelum dan sesudah iradiasi adalah Spectrophotometer Genesys 5 buatan Amerika. Setelah dosimeter CTA diiradiasi, disarankan pengukuran rapat optik/absorban ditunda (dikondisikan) dahulu selama ± 2 jam. Hal ini agar supaya diperoleh harga rapat optik/absorbanyang optimal, karena apabila pengukuran dilakukan sebelum 2 jam setelah diiradiasi, maka harga rapat optik/absorban di bawah nilai sebenarnya, dan jika pengukuran dilakukan setelah 2 jam maka harga rapat optik/absorbanakan bertambah besar seiring dengan bertambahnya waktu. Walaupun demikian pertambahan ini dapat dikoreksi dengan kurva hubungan antara Perubahan relatif rapat optik/absorban vs waktu setelah iradiasi seperti ditunjukkan pada Gambar 2. Untuk itu disarankan pengukuran rapat absorban dilakukan 2 jam setelah dosimeter diradiasi. Perhitungan dosis serap dilakukan berdasarkan Kurva kalibrasi dosis serap vs respon dosimeter. Kurva kalibrasi diperoleh dari hasil pengukuran absorban pada dosimeter yang telah diketahui nilai dosisnya secara pasti, dan hasilnya ditunjukkan pada Gambar 3. Sedangkan respon dosimeter (R) merupakan hasil pengukuran absorban dosimeter setelah diiradiasi (abs) dikurangi dengan absorban dosimeter sebelum diradiasi (bgd) dibagi dengan tebal dosimeter (t), jika ditulis secara metematik R = (abs bgd)/t. Setelah dosimeter CTA dikondisikan kurang lebih 2 jam, maka segera dilakukan pembacaan nilai rapat optik/absorban, pembacaan masing-masing dosimeter dilakukan sebanyak lima kali dan hasil perhitungan dosis serap ditunjukkan pada Tabel 1. Gambar 2. Perubahan relatif absorban vs waktu setelah iradiasi (4) Tabel 1. Hasil Pengukuran Dosimeter CTA Posisi Absorban Dosis (kgy) Dosis rerata (kgy) D maks /D min Selatan 0,218 13,4 11,8 14,8/11,8 = 1,25 0,199 10,6 0,221 13,8 0,203 11,2 0,193 9,7 Tengah 0,237 16,2 14,8 14,8/14,8 = 1 0,218 13,4 0,222 14,0 0,226 14,6 0,236 16,0 Utara 0,176 7,2 8,5 14,8/8,5 = 1,74 0,172 8,8 0,192 6,6 0,187 9,6 0,198 10,5 Sukaryono, dkk. ISSN 1410 8178 Buku II hal. 445

KESIMPULAN Gambar 3. Kurva kalibrasi dosis serap vs respon dosimeter CTA Dari hasil identifikasi arus berkas elektron pada MBE Lateks dapat disimpulkan sebagai berikut: Pada saat MBE Lateks dioperasikan pada tegangan pemercepat 208 kv dan arus berkas 25 µa, berkas elektron sudah sampai ke target yang ditandai dengan berubahnya warna dosimeter gonogo dari warna kuning menjadi warna merah kecoklatan. Distribusi arus berkas elektron belum merata (homogen) pada luasan window MBE, dosis tertinggi terdapat pada posisi tengah dengan dosis 14,8 kgy, kemudian posisi selatan dengan dosis 11,8 kgy dan dosis terendah pada posisi utara dengan dosis 8,5 kgy. Dari hasil pengukuran keseragaman dosis sepanjang window MBEdiperoleh D maks /D min = 1,74,dengan demikian keseragaman dosis sepanjang window MBE tidak homogen, hal ini dapat diperbaiki dengan menyempurnakan sistem optik yang belum bekerja secara maksimal. UCAPAN TERIMA KASIH Gambar 4. Kurva hubungan atara dosis versus posisi sepanjang window MBE Selanjutnya berdasarkan hasil perhitungan dosis serap dapat dibuat kurva hubungan antara posisi dosimeter CTA versus besar dosis, sehingga bisa dilihat distribusi berkas elektron yang mengenai target sepanjang window MBE. Berdasarkan data dan kurva dapat dilihat bahwa distribusi arus berkas elektron belum merata di sepanjang window MBE, dosis tertinggi terdapat pada posisi tengah dengan dosis 14,8 kgy, kemudian pada posisi selatan 11,8 kgy dan posisi utara 8,5 kgy. Dalam proses iradiasi, pengukuran dosis maksimal (D maks ) dan dosis minimal (D min ) sangat penting, dan bila terdapat perbedaan yang amat menyolok dari harga D maks dan D min berarti iradiasi tidak homogen. Homogenitas atau keseragaman distribusi dosis radiasi pada sepanjang window MBE dinyatakan dari perbandingan D maks /D min. Bila harga ini bernilai mendekati 1, maka distribusi dosis dapat dikatakan homogen. Dalam praktek khususnya dalam industri, perbandingan D maks /D min dapat mencapai 1 sampai 1,5 (5). Dari data pengukuran keseragaman dosis sepanjang window MBE, diperoleh keseragaman dosis serap D maks /D min = 1,74, dengan demikian keseragaman dosis sepanjang window MBE menunjukkan tidak homogen. Keseragaman distribusi dosis pada luasan window MBE masih bisa ditingkatkan dengan menyempurnakan sistem optik yang belum bekerja secara maksimal. Tiada kata yang lebih indah untuk diucapkan selain terima kasih, untuk itu penulis haturkan terimakasih yang setulus-tulusnya kepada Kapok Teknologi Akselerator, Kabid Teknologi Akselerator dan Fisika Nuklir serta semua tim MBE 300 kev/20 ma untuk pra vulkanisasi latek atas segala informasi dan data sehingga bisa tersusun makalah ini, semoga Allah SWT selalu memberikan imbalan yang setimpal atas jasa baik tersebut dan selalu memberi kekuatan kepada kita semua untuk berkarya dan beramal lebih baik lagi dalam setiap kesempatan. Amin. DAFTAR PUSTAKA 1. Rancangan Detil Mesin Berkas Elektron 300 kev/20 ma untuk Industri Lateks, Pusat Teknologi Akselerator dan Proses Bahan BATAN Yogyakarta (2006) 2. Laporan Analisis Keselamatan Konstruksi Mesin Berkas Elektron untuk Industri Lateks, BATAN Yogyakarta (2012) 3. RANY SAPTAAJI, Petunjuk Praktikum Dosimetri Akselerator Pelatihan Pekerja Akselerator, Pusat Pendidikan dan Pelatihan Badan Tenaga Nuklir Nasional Yogyakarta (2009). 4...., Sertikat CTA FTR-125, Fuji Film, Jepang. 5. MARGA UTAMA, Aplikasi Akselerator Untuk Industri, Diktat Pelatihan Pekerja Akselerator, Pusdiklat BATAN, Jakarta 2003. Buku II hal. 446 ISSN 1410 8178 Sukaryono, dkk

TANYA JAWAB Jumari Apakah dengan arus 25µA sudah cukup untuk operasi pemayaran MBE latek ini? Sukaryono Dengan arus berkas sebesar 25µA sudah bisa untuk mengidentifikasi distribusi berkas sepanjang window MBE. dilakuakan agar bisa mendapatkan hasil yang homogeny? Sukaryono Setelah diketahui bahwa keseragaman dosis sepanjang window tidak homogen langkah yang dilakukan adalah penyempurnaan terhadap system optik. Slamet wiranto Setelah diketahui keseragaman dosis sepanjang window MBE tidak homogen, langkah apa yang Sukaryono, dkk. ISSN 1410 8178 Buku II hal. 447