KONTROL KUALITAS TERAPI RADIASI PADA UNIT RADIOTERAPI MRCCC RS MRCCC

dokumen-dokumen yang mirip
Jurusan Fisika, Fakultas Matematika dan Ilmu Pengetahuan Alam, Universitas Hasanuddin

Analisis Dosis Keluaran Berkas Foton dan Elektron Energi Tinggi Pesawat Linac Elekta Precise 5991 Berdasarkan Code of Practice IAEA TRS 398

ANALISIS KUALITAS RADIASI DAN KALIBRASI LUARAN BERKAS FOTON 6 DAN 10 MV PESAWAT PEMERCEPAT LINIER MEDIK VARIAN CLINAC CX 4566 ABSTRAK

BAB II LINEAR ACCELERATOR

OPTIMASI ASPEK KESELAMATAN PADA KALIBRASI PESAWAT RADIOTERAPI

ANALISA KURVA PERCENTAGE DEPTH DOSE (PDD) DAN PROFILE DOSE UNTUK LAPANGAN RADIASI SIMETRI DAN ASIMETRI PADA LINEAR ACCELERATOR (LINAC) 6 DAN 10 MV

KENDALI KUALITAS DAN JAMINAN KUALITAS PESAWAT RADIOTERAPI BIDIKAN BARU LABORATORIUM METROLOGI RADIASI

ANALISIS POSISI DETEKTOR TERHADAP STEM EFFECT DAN DOSIS RELATIF UNTUK DOSIMETRI PESAWAT LINAC 6 MV

UNIVERSITAS INDONESIA VERIFIKASI PENYINARAN IMRT MENGGUNAKAN 2D ARRAY MATRIXX EVOLUTION SKRIPSI YAHYA MUSTOFA

Jumedi Marten Padang*, Syamsir Dewang**, Bidayatul Armynah***

PERBANDINGAN PENGUKURAN PDD DAN BEAM PROFILE ANTARA DETEKTOR IONISASI CHAMBER DAN GAFCHROMIC FILM PADA LAPANGAN 10 X 10 CM 2

BAB I PENDAHULUAN. kanker yang ditemukan di Indonesia (Riskesdas, 2007). Hal ini seiring dengan

Pengaruh Ketidakhomogenan Medium pada Radioterapi

BAB III PROTOKOL PENANGANAN KANKER PROSTAT DENGAN EKSTERNAL BEAM RADIATION THERAPY (EBRT)

Verifikasi Keluaran Radiasi Pesawat Linac (Foton Dan Elektron) Serta 60CO Dengan TLD

VERIFIKASI PENENTUAN LAJU DOSIS SERAP DI AIR BERKAS FOTON 6 MV DAN 10 MV PESAWAT PEMERCEPAT LINIER MEDIK CLINAC 2100 C MILIK RUMAH SAKIT

ANALISIS KUALITAS BERKAS RADIASI FOTON 10 MV PADA PESAWAT TELETERAPI LINEAR ACCELERATOR

PERAN FISIKAWAN MEDIS DALAM PEMANFAATAN TENAGA NUKLIR DI BIDANG KESEHATAN: RADIOTERAPI, RADIODIAGNOSTIK, KEDOKTERAN NUKLIR

Verifikasi TPS untuk Dosis Organ Kritis pada Perlakuan Radioterapi Area Pelvis dengan Sinar X 10 Megavolt

Desain dan Analisis Pengaruh Sudut Gantri Berkas Foton 4 MV Terhadap Distribusi Dosis Menggunakan Metode Monte Carlo EGSnrc Code System

PROFIL BERKAS SINAR X LAPANGAN SIMETRIS DAN ASIMETRIS PADA PESAWAT LINAC SIEMENS PRIMUS 2D PLUS

PENGARUH VARIASI AIR GAP TERHADAP DOSIS SERAP PENYINARAN BERKAS ELEKTRON PADA PESAWAT LINAC SIEMENS / PRIMUS M CLASS 5633

ANALISIS KARAKTERISTIK PROFIL PDD (PERCENTAGE DEPTH DOSE) BERKAS FOTON 6 MV DAN 10 MV

VERIFIKASI DOSIMETRI PERHITUNGAN BERKAS TERBUKA PERANGKAT LUNAK IN-HOUSE TREATMENT PLANNING SYSTEM (TPS) PESAWAT TELETERAPI COBALT-60

HUBUNGAN ANTARA LAJU DOSIS SERAP AIR DENGAN LAPANGAN RADIASI BERKAS ELEKTRON PESAWAT PEMERCEPAT LINIER MEDIK ELEKTA

ANALISA DOSIS RADIASI KANKER MAMMAE MENGGUNAKAN WEDGE DAN MULTILEAF COLLIMATOR PADA PESAWAT LINAC

ANALISIS PERHITUNGAN DOSIS SERAP TERAPI ROTASI DENGAN METODE TISSUE PHANTOM RATIO (TPR) PADA LINEAR ACCELERATOR (LINAC) 6 MV

OPTIMASI ASPEK KESELAMATAN PADA KALIBRASI PESAWAT TERAPI 60 Co atau 137 Cs

ANALISIS PROFIL BERKAS RADIASI LINEAR ACCELERATOR 6MV PADA PENGGUNAAN VIRTUAL WEDGE DENGAN GAFCHROMIC FILM

DESAIN DASAR PERANGKAT RADIOTERAPI EKSTERNAL MENGGUNAKAN COBALT-60

PENGARUH SUDUT GANTRI TERHADAP KONSTANSI DOSIS SERAP DI AIR PESAWAT TELETERAPI Co-60 XINHUA MILIK RUMAH SAKIT dr. SARJITO YOGYAKARTA

ANALISIS POSISI SUMBER RADIOAKTIF COBALT PADA PESAWAT TELETERAPI COBALT-60. Skripsi Untuk memenuhi sebagian persyaratan mencapai derajat Sarjana S-1

ANALISIS DOSIS OUTPUT SINAR-X PESAWAT LINEAR ACCELERATOR (LINAC) MENGGUNAKAN WATER PHANTOM

Jusmawang, Syamsir Dewang, Bidayatul Armynah Jurusan Fisika, Fakultas Matematika dan Ilmu Pengetahuan Alam, Universitas Hasanuddin

Verifikasi Distribusi Dosis Tps Dan Pesawat Linac Menggunakan Phantom Octavius 4d Dengan Teknik IMRT Protokol Kanker Lidah

Analisis Perubahan Kurva Percentage Depth Dose (PDD) dan Dose Profile untuk Radiasi Foton 6MV pada Fantom Thoraks

PENENTUAN DOSIS SERAP LAPANGAN RADIASI PERSEGI PANJANG BERKAS FOTON 10 MV DENGAN PENGUKURAN DAN PERHITUNGAN

ANALISIS DOSIS SERAP RADIASI PADA PERBEDAAN DIMENSI DAN BENTUK LAPANGAN PENYINARAN BERKAS RADIASI FOTON 6 MV

ANALISIS HASIL PENGUKURAN PERCENTAGE DEPTH DOSE (PDD) BERKAS ELEKTRON LINAC ELEKTA RSUP DR. SARDJITO

PENENTUAN PARAMETER DOSIMETRI AWAL BERKAS FOTON 6 MV DARI 5 BUAH PESAWAT PEMERCEPAT LINIER MEDIK ELEKTA DAN VARIAN CLINAC BARU

BAB I PENDAHULUAN. utama kematian akibat keganasan di dunia, kira-kira sepertiga dari seluruh kematian akibat

ANALISIS DOSIS OUTPUT BERKAS ELEKTRON PESAWAT TELETERAPI LINEAR ACCELERATOR (LINAC)TIPE VARIAN HCX 6540 MENGGUNAKAN TRS 398

ANALISIS KUALITAS CITRA VERIFIKASI LAPANGAN RADIASI LINAC PADA KANKER PAYUDARA MENGGUNAKAN VARIASI MONITOR UNIT. Skripsi FRILYANSEN GAJAH

FAKTOR KOREKSI SOLID WATER PHANTOM TERHADAP WATER PHANTOM PADA DOSIMETRI ABSOLUT BERKAS ELEKTRON PESAWAT LINAC

PENENTUAN FAKTOR KELUARAN BERKAS ELEKTRON LAPANGAN KECIL PADA PESAWAT LINEAR ACCELERATOR

ANALISIS DOSIS PADA PENGGUNAAN FILTER WEDGE MENGGUNAKAN DOSIMETER GAFCHROMIC EBT2 DAN GAFCHROMIC XR-RV3 UNTUK BERKAS FOTON 6 MV

PENGARUH BLOK INDIVIDUAL BERBAHAN CERROBEND PADA DISTRIBUSI DOSIS SERAP BERKAS FOTON 6 MV LINEAR ACCELERATOR (LINAC)

PENENTUAN FAKTOR KELUARAN BERKAS ELEKTRON LAPANGAN KECIL PADA PESAWAT LINEAR ACCELERATOR

BAB I PENDAHULUAN. 1.1 Latar Belakang dan Rumusan Masalah. Penggunaan radiasi dalam bidang kedokteran terus menunjukkan

PERBANDINGAN DOSIS RADIASI DI UDARA TERHADAP DOSIS RADIASI DI PERMUKAAN PHANTOM PADA PESAWAT CT-SCAN

Pendidikan dan Peran Fisikawan Medik dalam Pelayanan Kesehatan

Jurusan Fisika, Fakultas Matematika dan Ilmu Pengetahuan Alam, Universitas Hasanuddin

BAB I PENDAHULUAN 1.1 Latar Belakang Masalah

KARAKTERISASI DETEKTOR IN VIVO UNTUK DOSIMETRI RADIOTERAPI EKSTERNA

Verifikasi Dosis Radiasi Kanker Menggunakan TLD-100 pada Pasien Kanker Payudara dengan Penyinaran Open System

Analisis Pengaruh Sudut Penyinaran terhadap Dosis Permukaan Fantom Berkas Radiasi Gamma Co-60 pada Pesawat Radioterapi

Berkala Fisika ISSN : Vol. 16, No. 4, Oktober 2013, hal

STUDI AWAL UJI PERANGKAT KAMERA GAMMA DUAL HEAD MODEL PENCITRAAN PLANAR STATIK MENGGUNAKAN SUMBER RADIASI HIGH ENERGY IODIUM-131 (I 131 )

VERIFIKASI BERKAS ELEKTRON PESAWAT LINEAR ACCELERATOR (LINAC) DENGAN VARIASI ENERGI PADA WATER PHANTOM Raden Asrisal, Syamsir Dewang, Dahlang Tahir

KOREKSI KURVA ISODOSIS 2D UNTUK JARINGAN NONHOMOGEN MENGGUNAKAN METODE TAR (TISSUE AIR RATIO)

PEREKAYASAAN PERANGKAT RADIOTERAPI EKSTERNAL MENGGUNAKAN COBALT-60

KARAKTERISASI DOSIMETRI SUMBER BRAKITERAPI IR-192 MENGGUNAKAN METODE ABSOLUT

BAB IV PERBANDINGAN DATA DAN ANALISIS JUMLAH MONITOR UNIT OUTPUT SOFTWARE ISIS DENGAN OUTPUT SIMULASI MONTE CARLO

Youngster Physics Journal ISSN : Vol. 4, No. 1, Januari 2015, Hal

EVALUASI VERIFIKASI LAPANGAN PENYINARAN PADA KANKER PAYUDARA MENGGUNAKAN TEKNIK INTENSITY MODULATED RADIOTHERAPY DENGAN BERBAGAI FRAKSI

PENGARUH KEDALAMAN PADA OUTPUT FACTOR LAPANGAN KECIL DENGAN DETEKTOR THERMOLUMINESENSE DOSIMETER RODS DAN IONIZATION CHAMBER EXRADIN A16

EVALUASI TEBAL DINDING RUANGAN PESAWAT LINEAR ACCELERATOR (LINAC) SINAR-X DI INSTALASI RADIOTERAPI RUMAH SAKIT UNIVERSITAS HASANUDDIN

UNIVERSITAS INDONESIA

Analisis Dosis Radiasi Kanker Nasofaring Dengan Menggunakan Wedge Pada Pesawat Linear Accelerator (LINAC)

BAB I PENDAHULUAN. 1.1 Latar Belakang dan Rumusan Masalah. Kemajuan pesat pada bidang radiotherapy telah banyak memberikan

PENENTUAN KARAKTERISASI CERROBEND SEBAGAI WEDGE FILTER PADA PESAWAT TELETERAPI 60 Co

Simulasi Monte Carlo untuk Menentukan Rasio Deposisi Dosis pada a-si Epid dengan Deposisi Dosis pada Air

PENGARUH DIAMETER PHANTOM DAN TEBAL SLICE TERHADAP NILAI CTDI PADA PEMERIKSAAN MENGGUNAKAN CT-SCAN

PENGUKURAN DOSIS RADIASI PADA PASIEN PEMERIKSAAN PANORAMIK. Abdul Rahayuddin H INTISARI

ANALISIS PENGGUNAAN POLYDIMETHYL SILOXANE SEBAGAI BOLUS DALAM RADIOTERAPI MENGGUNAKAN ELEKTRON 8 MeV PADA LINAC

ANALISIS DOSIS SERAP RELATIF BERKAS ELEKTRON DENGAN VARIASI KETEBALAN BLOK CERROBEND PADA PESAWAT LINEAR ACCELERATOR

BAB 1 PENDAHULUAN. Salah satu bentuk pemanfaatan radiasi pengion adalah untuk terapi atau yang

Estimasi Dosis Efektif Out of Field dalam Radioterapi LINAC Varian ix pada Kanker Serviks

BAB I PENDAHULUAN. A. Latar Belakang Masalah. proporsi usia lanjut (WHO, 2005, pp. 8-9). Di Indonesia, data survei kesehatan

Verifikasi Ketepatan Hasil Perencanaan Nilai Dosis Radiasi Terhadap Penerimaan Dosis Radiasi Pada Pasien Kanker

KAJIAN TENTANG PERHITUNGAN JUMLAH MONITOR UNIT PADA SOFTWARE ISIS YANG DIPAKAI DI BAGIAN TREATMENT PLANNING

Berkala Fisika ISSN : Vol. 18, No. 1, Januari 2015, hal 1-8

UJI KESESUAIAN PESAWAT CT-SCAN MEREK PHILIPS BRILIANCE 6 DENGAN PERATURAN KEPALA BAPETEN NOMOR 9 TAHUN 2011

BAB III PERHITUNGAN JUMLAH MONITOR UNIT MENGGUNAKAN METODE MONTE CARLO

PENGUKURAN FAKTOR WEDGE PADA PESAWAT TELETERAPI COBALT-60 : PERKIRAAN DAN PEMODELAN DENGAN SOFTWARE MCNPX.

IMPLEMENTASI LINEAR ACCELERATOR DALAM PENANGANAN KASUS KANKER

Youngster Physics Journal ISSN : Vol. 2, No. 1, April 2013, Hal 27-34

BAB IV PERHITUNGAN DOSIS SERTA ANALISIS PENGARUH UKURAN MEDAN PAPARAN TERHADAP OUTPUT BERKAS FOTON

BAB I PENDAHULUAN I.1. Latar Belakang

PERSIAPAN & TERAPI RADIASI PASIEN DGN STS RSUD Dr. SOETOMO SURABAYA

BAB 1 PENDAHULUAN. radionuklida, pembedahan (surgery) maupun kemoterapi. Penggunaan radiasi

PERHITUNGAN EFISIENSI DAYA BERDASAR PROSEN- TASE KEDALAMAN DOSIS (PDD) PADA LINAC MEDIS RS DR. SARDJITO

PHARMACIST CREDENSIALS IN THE INDONESIAN NATIONAL ACCREDITATION STANDARD 2012 VERSION

PERANCANGAN RUANGAN RADIOTERAPI EKSTERNAL MENGGUNAKAN SUMBER Co-60

Prediction of 2D Isodose Curve on Arbitrary Field Size in Radiation Treatment Planning System (RTPS)

STUDI AWAL UJI PERANGKAT KAMERA GAMMA DUAL HEAD MODEL PENCITRAAN PLANAR (STATIK) MENGGUNAKAN SUMBER RADIASI MEDIUM ENERGY RADIUM-226 (Ra 226 )

PERANCANGAN RUANGAN RADIOTERAPI EKSTERNAL MENGGUNAKAN SUMBER Co-60

PERANCANGAN PERISAI RADIASI PADA KEPALA SUMBER UNTUK PESAWAT RADIOTERAPI EKSTERNAL MENGGUNAKAN CO-60 PADA POSISI BEAM OFF

BAB V ANALISIS KONSISTENSI PROTOKOL PENANGANAN RADIASI KANKER PROSTAT DENGAN EBRT PADA RS.X

Jurnal Fisika Unand Vol. 3, No. 3, Juli 2014 ISSN

Transkripsi:

KONTROL KUALITAS TERAPI RADIASI PADA UNIT RADIOTERAPI MRCCC Fielda Djuita 1, Rina Taurisia 2 & Andreas Nainggolan 2 1 Kepala Unit Radioterapi 2 Fisikawan Medis RS MRCCC ABSTRAK KONTROL KUALITAS TERAPI RADIASI PADA UNIT RADIOTERAPI MRCCC. Meningkatnya jumlah penderita kanker di Indonesia tidak sejalan dengan ketersediaan sarana terapi kanker, terutama sarana terapi radiasi. Oleh karena itu, beberapa Rumah Sakit swasta berpartisipasi dalam menyediakan sarana ini, salah satunya Rumah Sakit Mochtar Riady Comprehensive Cancer Center (MRCCC). Sebagai Rumah Sakit yang baru melaksanakan terapi radiasi maka penulis mencoba untuk menguraikan pelaksanaan kontrol kualitas yang dilaksanakan di Rumah Sakit ini. Tujuan: Sebagai kontrol kualitas praktik klinik pelayanan radioterapi. Metode: Diuraikan secara deskriptif seluruh proses kontrol kualitas yang dilakukan setiap hari, minggu dan bulanan. Hasil: Rata- rata perubahan keluaran setiap bulan < 1.008 %. Secara mekanik sangat stabil. Perubahan profil flattnes sinar photon bulanan < 0.01 % dan simetri sinar stabil. Kesimpulan : Perubahan rata-rata keluaran sinar photon dan electron jauh dibawah batas toleransi. Kata kunci: kontrol kualitas,praktik klinik, Linear accelerator, radioterapi. ABSTRACT RADIATION THERAPY QUALITY CONTROL IN MRCCC RADIOTHERAPY UNITS. Increasing cancer patients in Indonesia is not supported with the number of equipment that is able to treat cancer patients, especially in the radiation therapy field. Therefore, several private hospitals have joined to provide radiation therapy services and one of them is MRCCC. As a new hospital providing services in radiotherapy field, the writer tries to present our quality control program that we have done in our hospital. Purpose: As quality control to radiation therapy clinical practice. Methods: Descriptive essay of what we do in our institution. Conclusion: Average output photon and electron lower more than tolerance dose. Keywords: quality control, clinical practice, linear accelerator, radiotherapy. 1. Pendahuluan 1.1. Latar Belakang Departemen Kesehatan Republik Indonesia memperkirakan angka penderita kanker di Indonesia adalah 100 per 100.000 penduduk. Bila jumlah penduduk Indonesia adalah 230 juta, diperkirakan 230.000 kasus baru ditemukan setiap tahun. Hasil survei kesehatan rumah tangga menunjukkan bahwa proporsi kematian oleh penyakit menular menurun dari 69,49% (1980) menjadi 44,5% (2001), sedangkan kematian karena penyakit tidak menular termasuk kanker, meningkat dari 25,41% menjadi 48,53% pada tahun yang sama (Ditjen PPTM, 2006). Diperkirakan akan terjadi peningkatan dari 10 juta kasus kanker baru di dunia tahun 2000 menjadi 15 juta pada tahun

2020 (WHO, 2003). Hampir dua per tiga pasien-pasien kanker akan mendapat radiasi dalam pengobatannya. Di Amerika 75% pasien-pasien kanker mendapat radiasi sebagai terapi utama, 66% pasien kanker memerlukan radiasi selama masa sakitnya, 45% untuk terapi kuratif dan 21% untuk terapi paliatif (ASTRO, 2008). Bila dibuat perhitungan, maka penderita kanker baru di Indonesia yang memerlukan terapi radiasi lebih kurang 151.800 orang per tahun. Berdasarkan rekomendasi International Atomic Energy Agency agar setiap 500 pasien dapat dilayani oleh 1 pesawat radiasi (IAEA, 2007), diperlukan 303 pesawat linear accelerator atau cobalt. Sementara di Indonesia hanya ada 21 pusat radiasi dimana yang aktif hanya 20 pusat radiasi dengan 33 pesawat radiasi, sehingga kapasitas pelayanan kurang memadai. Data dari Perhimpunan Onkologi Radiasi Indonesia menyatakan bahwa hanya 14553 pasien yang terlayani per tahun jadi tidak sampai 10% (data PORI tahun 2010). Di Indonesia sekarang telah ada penyelenggaraan Pelayanan Radioterapi yang dilaksanakan oleh pihak swasta, salah satunya adalah Rumah Sakit Mochtar Riady Comprehensive Cancer Center (RS MRCCC). Sebagai Rumah Sakit baru, penulis menyampaikan kontrol kualitas di Rumah sakit ini dan mengharapkan masukan dari peserta seminar dalam kontrol kualitas. 1.2. Dasar Teori Kontrol Kualitas American Association of Physicists in Medicine TG 40 tahun 1994 dipakai secara luas untuk test jaminan mutu pesawat linear accelerator. Tetapi setelah itu berkembang beberapa teknologi baru pada praktik klinis seperti pemakaian Multi Leaf Collimator (MLC), asimetris jaw, wedge dinamik dan virtual, electronic portal imaging devices (EPIDs), conebeam CT, static kilovoltage (kv) dan respiratory gating yang dulu masih jarang digunakan. Dokumen TG 40 tidak dapat memenuhi kebutuhan tersebut diatas sehingga dibuatlah dokumen TG 142. Sejalan perkembangan TG 142, terjadi juga perkembangan teknik radiasi dengan penggunaan MLC pada IMRT sekaligus berputarnya gantri pesawat. Jaminan mutu teknik ini belum termasuk TG 142. Selanjutnya TG 142 menyatakan dokumen ini dapat sebagai guideline tetapi setiap institusi menyesuaikan dengan kebutuhan dan memodifikasi

sesuai kebutuhannya masing-masing (Kutcher 1994; Klein 2009). 2. Metodologi Pelaksanaan kontrol kualitas pada institusi kami terdiri dari harian, mingguan dan bulanan seperti dibawah ini: diperiksa setiap hari. Rangkuman kontrol kualitas harian dan mingguan dapat dilihat pada Tabel 1, di atas. Pengecekan ukuran kolimator bertujuan untuk memastikan kesesuaian antara bacaan luas lapangan kolimator yang sudah diatur pada kontrol panel atau pada layar monitor, bacaan skala mekanik, dan posisi sesungguhnya. (a) 2.1. Kontrol Kualitas Harian Tujuan dilakukannya kontrol kualitas harian adalah untuk memastikan bahwa parameter- parameter yang ada telah di setting dengan benar sehingga tidak akan terjadi perubahan yang dapat mempengaruhi dosis pada pasien. Oleh karena itu baik laser, optical distance indicator (ODI), dosis pasien (output constancy) dan keamanan (door interlock dan audiovisual contact) harus (b) Gambar 1. (a) Layar tampilan ukuran kolimator dan gantry pada Linac. (b) Pengecekan luas lapangan sinar kolimator dilihat pada kertas grafik

Pengecekan dan penyesuaian laser, bertujuan untuk mengetahui kesesuaian penanda laser yang berasal dari 4 arah berbeda dan memastikan serta menyesuaikan bahwa sinar laser presisi dan akurat di posisi tengah (isocenter) lapangan penyinaran. dengan target, serta memastikan bahwa posisi pasien atau target pada jarak tertentu dari sumber radiasi presisi dan akurat ketika menggunakan baik itu jarak SSD atau teknik isosentrik. (a) (a) (b) (b) Gambar 2. (a) Penyesuaian laser dengan menggunakan Wilke Phantom. (b) Penyesuaian laser terhadap cross hair gantry Pengecekan optical distance indicator (ODI) bertujuan untuk mengetahui kesesuaian jarak antara sumber radiasi Gambar 3. (a) Indikator Cahaya SSD (b) Komparasi Indikator Cahaya SSD dengan Indikator Jarak SSD Mekanik Pengecekan output pesawat Linac bertujuan untuk mengukur dosis keluaran linac setiap hari. Hal ini dapat dilakukan dengan 3 cara: (a) dengan

menggunakan bilik ionisasi tipe Farmer dan 1D Water Phantom, (b) dengan menggunakan bilik ionisasi tipe Farmer dan solid water phantom, atau (c) dengan menggunakan IBA Dosimetry StarTrack*. Cara (a) lebih direkomendasikan untuk pengecekan output mingguan karena lebih akurat dan presisi untuk pengukuran absolut. Sedangkan cara (b) dan (c) bisa digunakan untuk pengecekan output harian karena membutuhkan waktu yang relatif singkat dalam pengaturan perlengkapan dan pengambilan data. Gambar 4. Pengecekan Output Linac dengan bilik ionisasi tipe Farmer dan 1D Water Phantom Gambar 5. Pengecekan Output Linac dengan bilik ionisasi tipe Farmer dan Solid Water Phantom

indikator jarak meja pasien (baik lateral, longitudinal, ketinggian ataupun rotasi), luas lapangan cahaya penyinaran, beam flatness, profile consistency. Selain itu, dilakukan juga verifikasi pretreatment untuk setiap kasus IMRT dengan menggunakan IBA Dosimetry MatriXX Evolution (sebuah ion chambers array detector dengan luas lapangan maksimal 28x28 cm 2 ). Gambar 6. Pengecekan Output Linac dengan IBA Dosimetry StarTrack* 2.2. Kontrol Kualitas Bulanan Kontrol kualitas bulanan biasanya dilakukan pada parameter-parameter yang mengalami perubahan-perubahan secara minor (kecil), karena kemungkinan perubahan parameterparameter tersebut sangat kecil dalam waktu satu bulan. Kontrol kualitas bulanan ditujukan untuk parameter parameter yang mempunyai dampak kecil terhadap pasien, sebagai contoh: Gambar 7. Setup MatriXX Evolution untuk verifikasi IMRT Dikombinasikan dengan MultiCUBE

Output (cgy / 100 MU) 3. Hasil dan Pembahasan Hasil pengukuran harian seperti kurva 1. 6 MV Profile FS 10x10 cm (IBA Dosimetry StarTrack*) Daily Output Constancy Check May 2011 May 2011 June 2011 103.00 102.00 101.00 Gambar 10. Kurva bulanan dgn Start Track. 100.00 99.00 98.00 97.00 1 6 11 16 21 26 31 Gambar 8. Day of the Month Kurva harian selama bulan Mei, 2011 6 MV 15 MV 6 MV ref 15 MV ref Hasil perubahan rata-rata bervariasi tidak sampai 1%. Dibawah ini flatness dan simetris kurva bulanan. 6 MV Profile FS 10x10 cm @ 10 cm depth (IBA Dosimetry Blue Phantom 2 & CC13 Ion Chamber) Selanjutnya setiap penyinaran dengan tehnik IMRT harus dibuat verifikasi dengan MatriXX dikombinasikan dengan MultiCUBE. Dalam verifikasi untuk IMRT dilakukan 2 tahapan; tahap pertama yaitu menentukan output Linac yang di planning dengan real hasil penyinaran menggunakan open field 20x20, 237 MU = 200 cgy. Tahap kedua melakukan verifikasi hasil planning IMRT dengan hasil penyinaran verifikasi dosis IMRT menggunakan alat bantu MatrixXX evolution. Lihat gambar 11 dan 12 dibawah ini. Gambar 9. Kurva bulanan dgn blue phantom Tahapan Pertama Gambar planning dosis pesawat Linac pada lapangan terbuka (open field) 20x20 hasil planning dari TPS. Untuk mengetahui besarnya Output pesawat Linac yang diplanning dengan Output hasil penyinaran. 237 MU = 200 cgy Planning 237 MU = 199.3 cgy Hasil penyinaran. Gambar 11

Tahapan Kedua bulanan secara rutin, selain itu pemakaian MatriXX telah dilakukan sebelum terapi IMRT. Gambar planning dosis IMRT dari TPS 1x Fraksinasi 212,7 cgy Gambar 12 Kontrol kualitas digambarkan sebagai suatu program yang dirancang untuk melakukan kontrol dan memelihara standar kualitas. Dalam bidang onkologi radiasi, program jaminan mutu adalah suatu kebijakan dan prosedur yang penting dalam memelihara kualitas penatalaksanaan pasien. Menurut IAEA ada 3 komponen yang penting dalam membangun suatu pusat radioterapi yaitu: peralatan, sumber daya manusia dan prosedur. Agar dapat menjalankan suatu pusat Radioterapi secara efektif, efisien dan aman, perlu suatu peralatan yang memadai, sumber daya manusia yang memadai dan staf yang terlatih serta prosedur yang tertata rapi. Selain itu dikatakan juga bahwa alat canggih saja tidak dapat mengobati pasien kanker, tetapi dokter Onkologi Radiasi dan staf yang mendapat pelatihan secara professional yang mampu melakukannya (IAEA 2007). Pada institusi kami dilakukan kontrol kualitas secara harian, mingguan dan 4. Kesimpulan Demikianlah uraian pelaksanaan kontrol kualitas pada pusat radioterapi yang baru dibuka di Rumah Sakit Mochtar Riady Comprehensive Cancer Center. Pada uraian di atas penulis menguraikan pelaksanaan kontrol kualitas pada pesawat linear accelerator serta pemakaian MatriXX sebagai alat verifikasi pra-radiasi IMRT. Tentu saja masih banyak lagi prosedur kontrol kualitas yang belum diuraikan seperti kontrol kualitas untuk Treatment Planning System. 5. Daftar Pustaka [1] ASTRO (2008). Fast fact data of cancer. Diunduh dari http://www.astro.org. Pada tanggal 16 Maret 2008. [2] Aznar MC, Petersen PM, Logadattir A, Lindberg H et all,. (2010). Rotational Radiotheraphy for prostate cancer in clinical practice, Radiotherapy and Oncology 97, 480-484

[3] Dirjen PPTM (2006). Direktorat jenderal pengendalian penyakit dan penyehatan lingkungan, departemen kesehatan Republik Indonesia. Profil Direktorat Pengendalian Penyakit Tidak Menular. Diunduh dari http://www.global/cancer pada tanggal 9 Agustus 2008. [4] Guertin Timothy E,. (2002). Annual report IMRT: Targeting cancer. The radiation Oncology Departement in The Future. Varian Medical System. [5] IAEA (2007). Comprehensive Audits of Radiotherapy Practices: A tool for quality improvement. Quality assurance Team for Radiation Oncology (QUATRO), International Atomic Energy Agency, Vienna. [6] IAEA (2007). Setting Up a Radiotherapy Program: Clinical, Medical Physics, Radiation Protection and Safety Aspects. International Atomic Energy Agency, Vienna. [7] Kutcher Gerald J et all., (1994). Comprehensive QA for radiation oncology: Report of AAPM Radiation Therapy Committee Task Group 40. Med. Phys. 21 ($), April. [8] Klein Eric E., (2009). Task Group 142 report: Quality Assurance of Medical Accelerators. Med. Phys. 36 (9), September. [9] WHO, G. (2003). Quality and accreditation in health care services a global review evidence and information for policy for Department of Health. [10] Zankowsky Corey., (2011). Safety in Radiation Therapy. Varian s Perspective. Primary Investigator Training. Las Vegas April, 14.