HUBUNGAN TEGANGAN DAN CITRA RADIOGRAFI REAL TIME PADA PESAWAT SINAR-X RIGAKU RADIOFLEX-250EGS3

dokumen-dokumen yang mirip
HUBUNGAN TEGANGAN DAN CITRA RADIOGRAFI REAL TIME PADA PESAWAT SINAR-X RIGAKU RADIOFLEX-250EGS3

Pengembangan Fasilitas Radiografi Neutron, RN1, untuk Tomografi

UJI FUNGSI MODUL PENANGKAP CITRA SINAR-X BERBASIS LAYAR PENDAR

UNJUK KERJA PENCITRAAN PADA MODUL PENANGKAP CITRA SINAR- BERBASIS LAYAR PENDAR

Konsep Dasar Pengolahan Citra. Pertemuan ke-2 Boldson H. Situmorang, S.Kom., MMSI

KALIBRASI AKUISISI CITRA PESAWAT SINAR-X PORTABLE DIG 1100

PERANCANGAN RUANGAN RADIOGRAFI MEDIK DI SEKOLAH TINGGI TEKNIK NUKLIR

REFURBISHING PESAWAT SINAR-X DIAGNOSTIK EKS. LITBANG BATAN

RANCANGAN SISTEM CATU DAYA DAN RUMAH PENANGKAP CITRA PADA PESAWAT SINAR-X FLUOROSCOPY

PERANCANGAN DAN PEMBUATAN ALAT UJI PINHOLE DAN MULTIHOLE UNTUK PENGUKURAN DIMENSI FOCAL SPOT PESAWAT SINAR-X

RANCANG BANGUN TEGANGAN TINGGI DC DAN PEMBALIK PULSA PADA SISTEM PENCACAH NUKLIR DELAPAN DETEKTOR

Dhahryan 1, Much Azam 2 1) RSUD 2 )Laboratorium Fisika Atom dan Nuklir Jurusan Fisika UNDIP

OPTIMALISASI DOSIS RADIASI SINAR-X TERHADAP PROYEKSI PA (POSTERO-ANTERIOR) DAN LAT (LATERAL) PADA TEKNIK PEMERIKSAAN FOTO THORAX SKRIPSI

STANDAR NASIONAL INDONESIA (SNI) BIDANG NUKLIR

Jurnal Fisika Unand Vol. 3, No. 2, April 2014 ISSN

PERANGKAT PENANGKAP CITRA SINAR-X BERBASIS LAYAR PENDAR

BAB III METODOLOGI PENELITIAN

PENGUJIAN LINIERITAS KELUARAN PEMBANGKIT ARUS SINAR X MENGGUNAKAN STEPWEDGE SKRIPSI. Evi Yusita Nim

PENENTUAN NILAI TEBAL PARUH (HVL) PADA CITRA DIGITAL COMPUTED RADIOGRAPHY

PEREKAYASAAN PERANGKAT PENANGKAP CITRA SINAR-X BERBASIS LAYAR PENDAR

BAB I PENDAHULUAN. Radiodiagnostik merupakan tindakan medis yang memanfaatkan radiasi

PERANCANGAN PERANGKAT SINAR-X DIGITAL UNTUK DIAGNOSIS MEDIS

III. METODOLOGI PENELITIAN

RANCANG BANGUN SISTEM ANTARMUKA RATEMETER DENGAN PRINTER MENGGUNAKAN KOMPUTER DAN MIKROKONTROLER ATMEGA8535

STUDI AWAL UJI PERANGKAT KAMERA GAMMA DUAL HEAD MODEL PENCITRAAN PLANAR STATIK MENGGUNAKAN SUMBER RADIASI HIGH ENERGY IODIUM-131 (I 131 )

BAB III METODE PENELITIAN. Daerah penelitian secarageografisterletakpada107 o o BT

PEMERIKSAAN KUALITAS BOOM FOOT MENGGUNAKAN TEKNIK UJI TAK RUSAK

Konversi Paparan pada Perubahan kv Pesawat Sinar- X Rigaku-RF-250EGM

BAB I PENDAHULUAN 1.1 Latar Belakang

STUDI RADIOGRAFI MAKRO DENGAN VARIASI JARAK SUMBER SINAR-BAYANGAN (SID) DAN UKURAN FOKUS TERHADAP PEMBESARAN BAYANGAN

PERANCANGAN KONSUL UNTUK OPERATOR PADA PEREKAYASAAN PESAWAT SINAR-X MAMOGRAFI

PERHITUNGAN GERAK FLEKSIBELITAS SUMBER RADIASI ISOTOP IR 192 DI DALAM LUBANG TUBE PAD A PERANGKAT BRAKITERAPI UNTUK TERAPI KANKER S

PENGARUH TEGANGAN TABUNG (KV) TERHADAP KUALITAS CITRA RADIOGRAFI PESAWAT SINAR-X DIGITAL RADIOGRAPHY (DR) PADA PHANTOM ABDOMEN

MODUL PERANGKAT LUNAK AKUISISI CITRA DAN KENDALI MEJA PUTAR PROTOTIPE PERANGKAT RADIOSKOPI UNTUK INDUSTRI MANUFAKTUR

PENENTUAN KEMBALI KOMPOSISI KOMPOSIT KARET ALAM TIMBAL OKSIDA SEBAGAI PERISAI RADIASI SINAR-X SESUAI KETENTUAN BAPETEN

Rekayasa Bahan untuk Meningkatkan Daya Serap Terhadap Gelombang Elektromagnetik dengan Matode Deposisi Menggunakan Lucutan Korona

PERATURAN KEPALA BADAN PENGAWAS TENAGA NUKLIR NOMOR 9 TAHUN 2011 TENTANG UJI KESESUAIAN PESAWAT SINAR-X RADIOLOGI DIAGNOSTIK DAN INTERVENSIONAL

BAB 2 TINJAUAN PUSTAKA

PRIMA Volume 8, Nomor 1, Juni 2011 ISSN : DESAIN PINTU RUANG PESAWAT SINAR-X DARI BAHAN KOMPOSIT KARET ALAM TIMBAL OKSIDA

KALIBRASI MONITOR AREA DI REAKTOR KARTINI YOGY A KART A Agung Nugroho PTKMR - BATAN

PENGEMBANGAN KAMERA CCD PADA FASILITAS RADIOGRAFI NEUTRON, RN1, DI BATAN SERPONG

ANALISIS KUALITAS RADIOGRAFI PADA OBJEK BERGERAK DAN OBJEK TIDAK BERGERAK DENGAN MENGGUNAKAN VARIASI EKSPOSE SKRIPSI

DAFTAR ISI. HALAMAN JUDUL... i. PERNYATAAN BEBAS PLAGIARISME... ii. HALAMAN PENGESAHAN... iii. HALAMAN TUGAS... iv. HALAMAN PERSEMBAHAN...

STUDI AWAL UJI PERANGKAT KAMERA GAMMA DUAL HEAD MODEL PENCITRAAN PLANAR (STATIK) MENGGUNAKAN SUMBER RADIASI MEDIUM ENERGY RADIUM-226 (Ra 226 )

X-Ray Fluorescence Spectrometer (XRF)

DETEKTOR RADIASI. NANIK DWI NURHAYATI, S.Si, M.Si nanikdn.staff.uns.ac.id

PERBANDINGAN KARAKTERISTIK KELUARAN ANTARA PESAWAT SINAR-X TOSHIBA MODEL DRX-1824B DAN TOSHIBA MODEL DRX-1603B. Skripsi

Rekonstruksi Citra Tomografi Neutron Untuk Aplikasi Teknologi Nuklir

RANCANGAN SISTEM CATU DAYA DAN RUMAH PENANGKAP CITRA PADA PESAWAT SINAR-X FLUOROSCOPY

UJI KESESUAIAN PESAWAT CT-SCAN MEREK PHILIPS BRILIANCE 6 DENGAN PERATURAN KEPALA BAPETEN NOMOR 9 TAHUN 2011

PEREKAYASAAN PESAWAT SINAR-X FLUOROSCOPY BERBASIS LAYAR PENDAR

Analisa Kualitas Sinar-X Pada Variasi Ketebalan Filter Aluminium Terhadap Dosis Efektif

PENGARUH TEKANAN GAS ISIAN ARGON ALKOHOL TERHADAP KARAKTERISTIK DETEKTOR GEIGER-MÜLLER TIPE SIDE WINDOW CARI RISTIANI M

RANCANGAN SISTEM CATU DAY A DAN RUMAH PENANGKAP CITRA PAD A PESAWAT SINAR-X FLUOROSCOPY

ANALISIS PAPARAN RADIASI LINGKUNGAN RUANG RADIOLOGI DI RUMAH SAKIT DENGAN PROGRAM DELPHI

KAJIAN PENGARUH WARNA DAN JARAK LAMPU PENGAMAN TERHADAP HASIL RADIOGRAF

PENGARUH GRID(KISI) LINIER TERHADAP KETAJAMAN DAN DENSITAS GAMBAR FILM RONTGEN PADA PEMOTOAN SCHEDEL LATERAL

ALAT UKUR RADIASI. Badan Pengawas Tenaga Nuklir. Jl. MH Thamrin, No. 55, Jakarta Telepon : (021)

KAJIAN PENGARUH WARNA DAN JARAK LAMPU PENGAMAN TERHADAP HASIL RADIOGRAF

PENGARUH RADIASI HAMBUR TERHADAP KONTRAS RADIOGRAFI AKIBAT VARIASI KETEBALAN OBYEK DAN LUAS LAPANGAN PENYINARAN MUHAMMAD SYARIF BODDY

KARAKTERISASI KACA TIMBAL UNTUK PELINDUNG PENANGKAP CITRA SINAR-X

Youngster Physics Journal ISSN : Vol. 2, No. 1, April 2013, Hal 27-34

Pengaruh Perubahan Ketebalan Terhadap Tegangan Ambang Dan Waktu

PEMBUATAN LINEAR AMPLIFIER MENGGUNAKAN LM318 UNTUK SPEKTROMETRI GAMMA

UJI KELAYAKAN PESAWAT SINAR-X TERHADAP PROYEKSI PA (POSTERO-ANTERIOR) DAN LAT (LATERAL) PADA TEKNIK PEMERIKSAAN FOTO THORAX

ANALISIS KESELAMATAN PESAWAT SINAR-X DI INSTALASI RADIOLOGI RUMAH SAKIT UMUM DAERAH SLEMAN YOGYAKARTA

UNTUK MENGURANGI RESIKO RADIASI P ADA OPERATOR. Ferry Suyatno PRPN -BAT AN PEREKA YASAAN SISTEM PESA W AT SINAR-X FLUOROSCOPY

PERKIRAAN DOSIS PASIEN PADA PEMERIKSAAN DENGAN SINAR-X RADIOGRAFI UMUM. RUSMANTO

INOVASI ALAT PENGATUR CATU DAYA TEGANGAN TINGGI PADA PESAWAT SINAR-X DIAGNOSTIK

PENGARUH JARAK TABUNG SINAR-X DENGAN FILM TERHADAP KESESUAIAN BERKAS RADIASI PADA PESAWAT X-RAY SIMULATOR DI INSTALASI RADIOTERAPI RSUD DR

KAMERA PENDETEKSI GERAK MENGGUNAKAN MATLAB 7.1. Nugroho hary Mindiar,

PENJELASAN TEKNIS UJI PROFISIENSI PESAWAT SINAR-X TAHUN 2015

PEREKAYASAAN SISTEM DETEKSI PERANGKAT SCINTIGRAPHY MENGGUNAKAN PSPMT

Image Formation & Display

PERANCANGAN DAN REALISASI DINDING PRESENTASI INTERAKTIF DENGAN PENDETEKSIAN POSISI SINAR POINTER LASER SEBAGAI OPERATOR KURSOR MOUSE ABSTRAK

ANALISA PEREKAMAN DAN REKONSTRUKSI HOLOGRAFI DIGITAL MIKROSKOPIK PADA KACA PREPARAT MENGGUNAKAN METODE IN-LINE. Mahasiswa Jurusan Fisika 2

Jenis-jenis Monitor. Gambar 1. CRT

BAB 1 PENDAHULUAN. Universitas Sumatera Utara

Perancangan Keselamatan Ruangan Radiologi Pesawat Sinar-X Di PSTA BATAN Yogyakarta

KARAKTERISASI FLAT- PANEL DETECTOR UNTUK PESAWAT SINAR-X DIGITAL

OPERASI MESIN BERKAS ELEKTRON (MBE) PTAPB BATAN TIPE BA 350 kev / 10 ma

PERANCANGAN RUANG PENGUJIAN KEBOCORAN PESAWAT SINAR X RIGAKU 250 KV DI STTN BATAN YOGYAKARTA

ANALISIS KESELAMATAN RADIASI PADA LABORATORIUM SINAR-X INDUSTRI STTN BATAN YOGYAKARTA

OPTIMASI PENGUKURAN KEAKTIVAN RADIOISOTOP Cs-137 MENGGUNAKAN SPEKTROMETER GAMMA

DAFTAR KELENGKAPAN DOKUMEN YANG HARUS DILAMPIRKAN

Gambar 17. Tampilan Web Field Server

gambar cavities, tersembunyi gigi struktur (seperti gigi bungsu), dan tulang kerugian yang

Desain Ulang Shielding Ruangan Linear Accelerator (Linac) untuk Keselamatan Radiasi Di Gedung 14 PSTA-BATAN Yogyakarta

INOVASI ALAT PENGATUR CATU DAYA TEGANGAN TINGGI PADA PESAWAT SINAR-X DIAGNOSTIK

Fotografi 1 Dkv215. Bayu Widiantoro Progdi Desain Komunikasi Visual Fakultas Arsitektur dan Desain Universitas Katolik SOEGIJAPRANATA

ANALISA PENGARUH GRID RASIO DAN FAKTOR EKSPOSI TERHADAP GAMBARAN RADIOGRAFI PHANTOM THORAX

DATA DIGITAL BENDA LANGIT

PENGUKURAN DOSIS RADIASI RUANGAN RADIOLOGI II RUMAH SAKIT GIGI DAN MULUT (RSGM) BAITURRAHMAH PADANG MENGGUNAKAN SURVEYMETER UNFORS-XI

BAB II SISTEM PENENTU AXIS Z ZERO SETTER

BAB III KALIBRASI DAN VALIDASI SENSOR KAMERA UNTUK PENGEMBANGAN RUMUS POSISI TIGA DIMENSI OBYEK

STMIK AMIKOM PURWOKERTO PENGOLAHAN CITRA. Akuisisi dan Model ABDUL AZIS, M.KOM

UJI LINE SCAN CAMERA PADA RANCANG BANGUN SISTEM PENCITRAAN PETI KEMAS DENGAN TEKNIK SERAPAN SINAR GAMMA

PENGUKURAN BENTUK PROFIL BERKAS ELEKTRON DARI SUMBER ELEKTRON TIPE PIERCE MENGGUNAKAN SENSOR TABUNG TV BEKAS

III. METODE PENELITIAN. Penelitian ini mulai dilaksanakan pada bulan Juli 2014 sampai Februari 2015.

PORTABLE DENSITOMETER BERBASIS PC VIA BLUETOOTH

Transkripsi:

SDM TEKNOLOGI SEMINAR NASIONAL NUKLIR VII YOGYAKARTA, 16 NOVEMBER 2011 HUBUNGAN TEGANGAN DAN CITRA RADIOGRAFI REAL TIME PADA PESAWAT SINAR-X RIGAKU RADIOFLEX-250EGS3 Zaenal Abidin, Muhamad Isa, Tri Wulan Tjiptono* zaenala6@gmail.com STTN-BATAN, *) PTAPB BATAN Yogyakarta JI. Babarsari Kotak Pos 6101NKBB Yogyakarta 55281 Telp: (0274) 484085, 489716, Fax: (0274) 489715 INTISARI HUBUNGAN TEGANGAN DAN CITRA RADIOGRAFI REAL TIME PADA PESAWAT SINAR-X RIGAKU RADIOFLEX-250EGS3. Telah dilakukan penelitian hubungan antara tegangan dan citra radiografi real time pesawat sinar-x rigaku radiojlex-250egs3. Penelitian dilakukan dengan menyinari lempeng sintillator BC 704 dengan pesawat sinar-x pada tegangan 110 kv s.d 250 kv dengan interval 10 kv danjarak antara pesawat sinar-x dengan sintilator adalah 50 cm, 55 cm, 60 cm, 65 cm dan 70 cm. Kelipan sintilator ditangkap dengan kamera ICCD yang dihubungkan dengan komputer. Citra yang diperoleh ditentukan nilai luminositasnya, kemudian dicari hubungan antara tegangan dengan luminisitas citra. Hasil penelitian menunjukkan bahwa tegangan operasi pesawat sinal' X menghasilkan citra dengan nilai luminitas tertentu, semakin tinggi tegangan, nilai luminitas dari citra yang dihasilkan semakin besar. Jarak semakin jauh, nilai luminositasnya semakin berkurang. Hubungan antara nilai luminisitas (y) dan tegangan (x) yang paling linear adalah y = 1,506x-155,8 dengan nilai R2 = 0,993 terletak padajarak 70 cm. Kata kunci : kamera ICCD, nilai luminositas, variasi tegangan dan jarak ABSTRACT RELATIONSHIP BETWEEN VOLTAGE AND REAL-TIME RADIOGRAPHIC IMAGE FROM RIGAKU RADIOFLEX -250EGS3 X-RAY MACHINE. The relationship between voltage and real-time radiographic image from the X-ray machine Rigaku radiojlex-250egs3 has been studied. The study was conducted by exposing X-rays to sintillator plate BC704 at 110 kv up to 250 kv at 10 kv interval and the distance between the X-ray machine with the scintillator is 50 cm, 55 cm, 60 cm, 65 cm and 70 cm. Light from the scintillator was captured by 1CCD camera connected to a computer. The image obtained is determined value of luminositas, then lookfor the relationship between the voltage to the image. The results showed that the operating voltage, the resulting image has a greater value luminosity. The distance the farther it will decrease the value of luminosity. The relationship between the luminosity (y) and voltage (x) is the most linear (y = 1.506 x-155,8) with the value of R2 = 0.993 located at a distance of 70 cm. Keywords: ICCD camera, the value of luminosity, voltage variation and distance 1. PENDAHULUAN Pengawasan penggunaan pesawat sinar-x harus dilakukan secara teratur dan berkelanjutan agar terjamin keselamatan pekerja, pasien dan masyarakat. Salah satu aspek yang diawasi adalah kebocoran pesawat sinar-x dan kemampuan penyinaran. Uji kebocoran biasa dilakukan dengan mengoperasikan pesawat pada kv dan ma maksimal, window tertutup, dan diukur paparan radiasinya pada jarak 1 m dari focal spopj. Kemampuan penyinaran sinar-x biasa dilihat dan kemampuan menghitamkan film, hal ini memerlukan waktu, untuk menghemat waktu maka dibutuhkan suatu alat yang dapat langsung menunjukkan kemampuan penyinaran pesawat sinar-x, alat ukur daya penyinaran merupakan alat yang dapat mengukur daya keluaran sinar-x, tetapi alat ini belum dimiliki. Pengukuran kemampuan penyinaran pesawat sinar-x dapat diketahui dari kemampuan menghasilkan citra real time yang dihasilkan dengan melihat nilai luminisitasnya. Data tentang nilai luminisitas suatu citra yang dihasilkan dan pengoperasian pesawat sinar-x belum banyak dipublikasikan, data ini penting karena dapat digunakan untuk mengukur kemampuan penyinaran pesawat sinar-x. Penelitian ini akan menentukan nilai luminisitas dari citra real time yang hasilkan Sekolah Tinggi Teknologi Nuklir-BATAN 400 Zainal Abidin dkk

SEMINAR NASIONAL SDM TEKNOLOGI NUKLIR VII YOGY AKARTA, 16 NOVEMBER 2011 dari penyinaran pesawat sinar-x pada tegangan 110 kv sampai dengan 250 kv dengan rentang 10 kv, nilai luminisitas ini dibandingkan dengan tegangan operasinya, kemudian dicari persamaan hubungannya. Hasil penelitian dapat dikembangkan untuk menentukan kemampuan daya penyinaran pesawat sinar-x, kalibrasi penunjuk tegangan pada pesawat sinar-x. 2. TEORI Sinar-X Proses terjadinya sinar-x dapat dibedakan berdasarkan kejadiannya, yaitu: 1. Radiasi sinar-x yang dihasilkan akibat perlambatan berkas elektron cepat mengenai anoda yang disebut pengereman laju elektron atau efek bremsstrahlung 2. Radiasi sinar-x yang dihasilkan akibat perpindahan elektron dari kulit fuar ke kulit lebih dalam yang disebut dengan sinar-x karakteristik. Pada dasamya pesawat sinar-x terdiri dari sumber elektron, tabung sinar-x, sumber tegangan tinggi yang mencatu tegangan listrik pada kedua elektroda dalam tabung sinar-x, dan unit pengatur. Pesawat sinar-x akan menjadi sumber radiasi apabila dioperasikan[2][3]. Layar Pendar (Detektor Sintilator BC 704). Detektor sintilator BC-704 adalah layar fosfor yang terbuat dari ZnS(Ag) dan 6Li. Detektor BC-704 biasanya berbentuk persegi dan datar. Layar BC-704 tebal standamya kira-kira 1 mm tebal pelat aluminum, tetapi dapat disesuaikan dengan pesanan ataupun permintaan[4]. Detektor ini berfungsi mengubah foton menjadi kelipan cahaya. Prosedur Penelitian 1. Pemasangan alat 2. Setting pencahayaan 3. Aging pesawat sinar X 4. Pengambilan data pada kv dan jarak tertentu 5.. Menyimpan citra video yang dihasilkan 6. Mengolah citra yang dihasilkan penuh, mampu menangkap banyangan sekecilpun yang terdiri generator penunda, unit gerbang tegangan tinggi dan unit kamera. Kamera ICCD (1lltellsifCharge Couplet! Device) Kamera ICCD adalah kamera yang dilengkapi dengan sensor CCD yang terintegrasi CCD (charge-coupled device) berfungsi mengubah cahaya menjadi elektron, cara kerja sensor ini seperti prinsip kerja sel surya. Sensor seperti memiliki ribuan bahkan jutaan sel surya yang kecil dalam bentuk matrik dua dimensi. Masing-masing sel akan mengubah cahaya dari sebagian kecil gambar yang ditangkap menjadi elektron. Langkah berikut adalah membaca nilai dari setiap sel di dalam gambar. Dalam kamera CCD, nilai tersebut dikirimkan ke dalam sebuah chip dan sebuah konverter analog ke digital mengubah setiap nilai piksel menjadi nilai digital [5J. 3. METODE PENELITIAN Alat dan Bahan Alat dan bahan yang adalah pesawat sinar-x rigaku radioflek 250EG S3, kamera ICCD, kontrol kamera iccd, detektor sintilator bc 704, laptop, kotak hitam, lensa kamera, cermin yang dilapisi kaca Pb dan tv tuner illtemal[6] Kont'rol Gambar 1. Skema alat penelitian 4. HASIL PEMBAHASAN Citra Sintilasi Dengan Variasi Tegangan Dalam penelitian Inl dilakukan dengan memvariasikan tegangan pesawat sinar-x dari 110 kv s.d 250 kv. Untuk citra real time radiografi yang dihasilkan dengan memvariasikan tegangan, data yang terjadi berupa data visual berbentuk video. Data visual berupa video diolah menjadi data digital dengan cara merubahnya menjadi gambar. Dengan bantuan program computer data digitalnya dapat diketahui. Data digitalnya adalah berupa nilai luminositas (nilai yang menunjukkan tingkat keputihan gambar). Tabel 1 adalah data digital yang diolah Zainal Abidin dkk 401 Sekolah Tinggi Teknologi Nuklir-BATAN

Contoh eitra radiografi real time ini dapat dilihat pad a Gambar 2. Nilai luminositas Gambar 2 dapat dilihat pada kolom Citra I tegangan 130 kv. Citra Sintilasi Dengan Variasi Jarak Variasi tegangan diberikan dengan interval 5 em, mulai dari jarak 50 em sampai dengan 70 em. Untuk data pada jarak 50 em dan selanjutnya sampai jarak 70 em gain menjadi tetap, yaitu 300. Tabel 2 merupakan data eitra luminositas pada jarak 50 em sampai dengan jarak 70 em dengan gain 300 yang sudah diolah menjadi data digital. Gambar 2. Citra pada tegangan 130 kv dengan jarak 50 em Tegangan 214,350 249,874245,858 237,944157,266 248,964242,988 242,716186,446 248,664239,470 246,208224,814 174,436132,436 253,546250,512 253,890250,666 250,688248,292 250,762249,346 165,320111,818 251,082250,082 253,660250,626 jarak Tabel 41,332 56,656 239,212 127,996 30,580 104,034 39,984 112,226 203,178 jarak 131,860 116,238 228,936 229,528 240,438 239,658 141,204 230,096 240,124 22,638 116,368 50,408 210,574 142,368 63,748 244,778 246,282 249,172 226,740 247,472 211,252 191,818 168,738 148,020 95,442 157,308 174,086 195,016 144,112 67,628 77,234 188,256 162,876 68,964 89,392 98,004 78,966 em 260 65 55 Histogram em luminositasnilai jarak luminositas 50 em s.d 70 em, gain 300 Dari Tabel 2 dibuat sebuah kurva hubungan tegangan pesawat sinar-x dengan eitra luminositas pada jarak 50 sampai dengan jarak 70 em dengan nilai luminositas di sumbu Y dan tegangan di sumbu X, tiap garisnya dibuat persamaan liniemya. Pada Gambar 3 terlihat bahwa garis yang paling linier berasal dari garis jarak 70 em dengan persamaan y = I,506x-155,8 dan nilai R~ = 0,993 Sekolah Tinggi Teknologi Nuklir-BA TAN 402 Zainal Abidin dkk

SEMINAR NASIONAL SDM TEKNOLOGI NUKLIR VII YOGY AKART A, 16 NOVEMBER 2011 300.00 ~ 0:3.t!... 150.00 250.00 VI 'E c>- 200.00 100.00 :;- 50.00 0.00 o 50 100 150 200 250 300 Tegangan (kv) --jarak 50 em --jarak 55 em --jarak 60 em --jarak 55 em --jarak 70 em Gambar 3. Hubungan luminositas terhadap tegangan sinar-x Dari data di atas dapat ditentukan bahwa semakin jauh jarak layar pendar maka puneak nilai luminositas akan menurun. Hal ini memang dikaitkan dengan nilai intensitas sinar-x yang diterima oleh layar pendar untuk berubah menjadi eahaya tampak, semakin jauh jarak layar pendar maka nilai intensitas sinar-x akan menurun. Dengan menurunnya nilai intensitas sinar-x maka eahaya tampak yang dihasilkan oleh layar pendar menjadi sedikit. Citra yang diperoleh dalam pengujian ini tidak berhubungan dengan kuadrat jarak. Hal ini dapat kita lihat dengan pendekatan nilai luminositas pada jarak 50 em sebagai nilai intensitas yang pertama, nilai luminositas pada jarak 55 em sebagai nilai intensitas yang kedua, nilai \uminositas pada jarak 60 em sebagai nilai intensitas yang ketiga, nilai luminositas pad a jarak 65 em sebagai nilai intensitas yang keempat, dan nilai luminositas pad a jarak 70 em sebagai nilai intensitas yang kelima. Diketahui bahwa interval jarak 5 em, setelah dihitung dengan rumusan Ie/I = r/i/ dan kita perbandingkan dengan data pad a Tabel 2, temyata hubungan kenaikan nilai luminositas dengan kuadrat jarak tidak ada hubungannya. Data untuk nilai luminositas yang seharusnya muneul dengan perhitungan Ie/I = r/i r2 dengan asumsi nilai awal pada jarak 50 em dapat terlihat pada Tabel 3. Zainal Abidin dkk 403 Sekolah Tinggi Teknologi Nuklir-BATAN

kuadrat jarak. SDM TEKNOLOGI SEMINAR NASIONAL NUKLIR VII YOGY AKARTA, 16 NOVEMBER 2011 Tabel 3. Nilai luminositas yang memakai perbandingan kuadrat jarak Tegangan jarak 253,55 214,35 250,76 237,94 248,96 250,69 248,66 242,72 249,87 246,21 253,89 253,66 165,32 174,36 251,08 jarak 148,80 103,33 127,03 109,24 148,56 121,27 126,89 148,08 143,84 145,91 210,73 150,46 129,40 210,54 150,32 129,28 137,22 115,26 84,26 210,44 176,77 150,26 129,22 127,96 144,72 121,56 177,91 149,44 208,40 208,13 148,61 127,80 127,76 165,89 141,01 206,64 173,58 147,54 208,07 206,39 173,37 147,36 126,73 127,35 201,45 169,22 123,70 207,40 125,48 204,35 177,01 176,85 175,05 174,83 197,49 174,78 174,21 171,66 97,97 88,86 em 70 55 60 65 em Nilai luminositas 5. KESIMPULAN Citra dapat diperoleh dari sintilator BC 704 yang disinari pesawat sinar X, dan diperoleh hubungan antara tegangan (kv) pesawat sinar-x dengan nilai luminitas citra yang dihasilkan. Semakin tinggi tegangan (kv) pesawat sinar-x, maka nilai luminitas citranya semakin besar. Jarak antara sintilator dengan pesawat sinar-x semakin jauh maka nilai luminitas citranya semakin berkurang. Hubungan antara nilai luminitas (y) dan tegangan pesawat sinar-x (x) yang paling optimum (linear) secara empiris dinyatakan dengan persamaan y = 1,506x-155,8 dengan nilai R2 = 0,993 terletak pada jarak 70 em. 5. James R. Janesick, 2001, Scientific chargecoupled devices, SPIE Press. Copyright. 6. Isa. M., 20 II, "Hubungan Tegangan dengan Citra Radiografi Real Time Pesawat Sinar X Rigaku Radioflex 250EGS3", STTN-BAT AN, Yogyakarta 6. DAFTAR PUSTAKA 1. Abidin, Z., 2010, "Nuklir dan Aplikasinya", STTN BAT AN, Yogyakarta 2. Wardhana, W A., 2006, "Teknologi Nuklir Proteksi Radiasi dan Aplikasiny", Penerbit ANDI, Yogyakarta 3. Panitia PPR STTN, 2010, Modul Pelatihan Petugas Proteksi Radiasi, STTN-BATAN, Yogjakarta 4. Anonim (2010). http://id.wikipedia.org/wiki/citra diakses pada tanggal 4 Agustus 2011. Sekolah Tinggi Teknologi Nuklir-BATAN 404 Zainal Abidin dkk