PENENTUAN CALIBRATOR SETTING CAPINTEC CRC-7BT UNTUK SAMARIUM-153

dokumen-dokumen yang mirip
Pengembangan Metode Pengukuran Aktivitas 226 Ra, 137 Cs dan 60 Co menggunakan Pencacah Kamar Pengion Detektor Merlin Gerin CPGB 1

PENENTUAN KARAKTERISTIK SERAPAN DAN KOEFISIEN ATENUASI LINIER PENYANGGA MYLAR TERHADAP RADIASI UNTUK SUMBER STANDAR Sr-90

PENENTUAN CALIBRATION SETTING DOSE CALIBRATOR CAPINTEC CRC-7BT UNTUK Ce-139

PENGUNGKUNGAN SUMBER 85 Kr, 133 Xe, 198 Au, DAN 24 Na PASCA IRADIASI

FAKTOR KOREKSI PENGUKURAN AKTIVITAS RADIOFARMAKA I-131 PADA WADAH VIAL GELAS TERHADAP AMPUL STANDAR PTKMR-BATAN MENGGUNAKAN DOSE CALIBRATOR

METODE KALIBRASI MONITOR GAS MULIA MENGGUNAKAN SISTEM SUMBER GAS KRIPTON-85 STATIS

INTERKOMPARASI PENGUKURAN KAPSUL DALAM Ir-192 UNTUK UJI TAK MERUSAK

Pemanfaatan Alat Standar Primer untuk Peningkatan Kualitas Nilai Kalibrasi Alat Ukur Radioaktivitas di Bidang Kedokteran Nuklir

PROGRAM JAMINAN KUALITAS PADA PENGUKURAN. RADIONUKLIDA PEMANCAR GAMMA ENERGI RENDAH:RADIONUKLIDA Pb-210

KAJIAN WAKTU PARO 90 Sr MENGGUNAKAN ALAT UKUR STANDAR DOSIMETER FARMER NE 2570/B DAN NE2570/A

ABSTRAK. PENDAHULUAN hasil produksi, teknologi nuklir dapat ABSTRACT

STANDARDISASI F-18 MENGGUNAKAN METODE SPEKTROMETRI GAMMA

OPTIMASI ASPEK KESELAMATAN PADA KALIBRASI PESAWAT TERAPI 60 Co atau 137 Cs

Widyanuklida, Vol. 15 No. 1, November 2015: ISSN

PENERAPAN KOEFISIEN KOREKSI ELEMEN KARTU THERMOLUMINISENCE (TLD) UNTUK PERHITUNGAN DOSIS EKSTERNA

UJI BANDING SISTEM SPEKTROMETER GAMMA DENGAN METODA ANALISIS SUMBER Eu-152. Nugraha Luhur, Kadarusmanto, Subiharto

OPTIMASI ASPEK KESELAMATAN PADA KALIBRASI PESAWAT RADIOTERAPI

Analisis Peluruhan Flourine-18 menggunakan Sistem Pencacah Kamar Pengion Capintec CRC-7BT S/N 71742

Metode Penentuan Nilai Kemampuan Ukur Terbaik (KUT) pada Perangkat Spektrometer Gamma

PENENTUAN AKTIVITAS SUMBER RADIOAKTIF PEMANCAR GAMMA Eu-152 DI LABORATORIUM PTNBR

EVALUASI DOSIS RADIASI EKSTERNAL PEKERJA PUSAT RADIOISOTOP DAN RADIOFARMAKA ( PRR )

EVALUASI DOSIS RADIASI EKSTERNAL PEKERJA PUSAT RADIOISOTOP DAN RADIOFARMAKA ( PRR )

DENGAN RAHMAT TUHAN YANG MAHA ESA KEPALA BADAN PENGAWAS TENAGA NUKLIR,

METODE STANDARDISASI SUMBER 60 Co BENTUK TITIK DAN VOLUME MENGGUNAKAN METODE ABSOLUT PUNCAK JUMLAH

Jurnal Fisika Unand Vol. 3, No. 3, Juli 2014 ISSN

Penentuan Aktivitas Radionuklida 60 Co Dengan Metode Ekstrapolasi Effisiensi

PEMANTAUAN DOSIS RADIASI INTERNAL DENGAN WBC UNTUK PEKERJA PUSAT TEKNOLOGI LIMBAH RADIOAKTIF SERPONG TAHUN 2012

INTERKOMPARASI PENGUKURAN OUTPUT IRADIATOR 137 Cs DAN PERSONAL DOSE EQUIVALENT, Hp(10) MENGGUNAKAN TLD DAN FILM

KALffiRASI ALA T UKUR AKTIVITAS DOSE CALIBRA TOR SECARA SIMUL TAN.

bahwa semakin besar jarak ukur maka dosis serap yang diterima semakin kecil. Kata kunci :Kalibrasi, survei meter, dosis serap, faktor kalibrasi

AUDIT MUTU PENGUKURAN DOSIS SERAP DARI SUMBER TELETERAPI Co-60 CIRUS 90131

PENENTUAN AKTIVITAS TRITIUM DAN KARBON-14 DENGAN METODA PENGUKURAN DUAL LABEL ABSTRACT

KETIDAKPASTIAN PENGUKURAN KEKASARAN PERMUKAAN KELONGSONG BAHAN BAKAR NUKLIR DENGAN ROUGHNESS TESTER SURTRONIC-25

KALIBRASI MONITOR AREA DI REAKTOR KARTINI YOGY A KART A Agung Nugroho PTKMR - BATAN

Jurnal Fisika Unand Vol. 3, No. 2, April 2014 ISSN

ANALISIS WAKTU PELURUHAN TERHADAP PERSYARATAN DOSIS RADIOISOTOP UNTUK PEMERIKSAAN GONDOK

ANALISIS KUALITAS RADIASI DAN KALIBRASI LUARAN BERKAS FOTON 6 DAN 10 MV PESAWAT PEMERCEPAT LINIER MEDIK VARIAN CLINAC CX 4566 ABSTRAK

ANTARBANDING PENGUKURAN AKTIVITAS ISOTOP 57 Co DAN 131 I (II)

OPTIMASI PENGUKURAN KEAKTIVAN RADIOISOTOP Cs-137 MENGGUNAKAN SPEKTROMETER GAMMA

PENGUKURAN KONSENTRASI RADON DALAM TEMPAT PENYIMPANAN LIMBAH RADIOAKTIF. Untara, M. Cecep CH, Mahmudin, Sudiyati Pusat Teknologi Limbah Radioaktif

PROGRAM KOMPUTER UNTUK MENGHITUNG AKTIVITAS CUPLIKAN SISTEM PENCACAH KAMAR PENGION MERLIN GERIN

PROGRAM KOMPUTER UNTUK MENGHITUNG AKTI- VITAS CUPLIKAN SISTEM PENCACAH KAMAR PENGION MERLIN GERIN

STUDI AWAL UJI PERANGKAT KAMERA GAMMA DUAL HEAD MODEL PENCITRAAN PLANAR STATIK MENGGUNAKAN SUMBER RADIASI HIGH ENERGY IODIUM-131 (I 131 )

PENENTUAN WAKTU TUNDA PADA KONDISIONING LIMBAH HASIL PENGUJIAN BAHAN BAKAR PASCA IRADIASI DARI INSTALASI RADIOMETALURGI

PEMANTAUAN LINGKUNGAN DI SEKITAR PUSAT PENELITIAN TENAGA NUKLIR SERPONG DALAM RADIUS 5 KM TAHUN 2005

STUDI AWAL UJI PERANGKAT KAMERA GAMMA DUAL HEAD MODEL PENCITRAAN PLANAR (STATIK) MENGGUNAKAN SUMBER RADIASI MEDIUM ENERGY RADIUM-226 (Ra 226 )

IMPLEMENTASI COMPLIANCE TEST PESAWAT DENTAL INTRAORAL PADA SALAH SATU KLINIK GIGI DI KOTA PADANG

PENGANGKUTAN LIMBAH RADIOAKTIF PADAT DAN CAIR DARI PENIMBUL KE INSTALASI PENGOLAHAN LIMBAH RADIOAKTIF. Arifin Pusat Teknologi Limbah Radioaktif -BATAN

PENGARUH KONTRIBUSI KETIDAKPASTIAN TERHADAP PELAPORAN NILAI POROSITAS MENGGUNAKAN METODE GRAVIMETRI

DOSIMETER CAS04:DY BUATAN BARC SEBAGAI PEMANTAU DOSIS RADIASI PERORANGAN HP (10) Rofiq Syaifudin, Nina Herlina, dan Bambang Supriyanto PTKMR - BAT AN

ESTIMASI KETIDAKPASTIAN PENGUKURAN DOSIS PERORANGAN MENGGUNAKAN THERMOLUMINISENCE DOSIMETER (TLD)

KALIBRASI EFISIENSI α/β COUNTER UNTUK ANALISIS RADIONUKLIDA PEMANCAR BETA DALAM CONTOH URIN

BAB I PENDAHULUAN. alat ukur suhu yang berupa termometer digital.

RADIOKALORIMETRI. Rohadi Awaludin

EVALUASI KINERJA SPEKTROMETER GAMMA YANG MENGGUNAKAN NITROGEN CAIR SEBAGAI PENDINGIN DETEKTOR

KOMPARASI PENGUKURAN LAJU KERMA UDARA PESAWAT OB-85 MENGGUNAKAN ALAT UKUR RADIASI STANDAR SEKUNDER DAN STANDAR TURUNANNYA

Jurnal Radioisotop dan Radiofarmaka ISSN Journal of Radioisotope and Radiopharmaceuticals Vol 9, Oktoberl 2006

OPTIMASI ALAT CACAH WBC ACCUSCAN-II UNTUK PENCACAHAN CONTOH URIN

PENENTUAN NILAI KETIDAKPASTIAN HASIL KALIBRASI DRYER OVEN MESIN SKRIPSI. Oleh: ARIE MULYA NUGRAHA

PENENTUAN NILAI ACUAN UJI BANDING ANTAR LABORATORIUM KALIBRASI UNTUK KALIBRASI MIKROPIPET BERDASARKAN KONSENSUS

Analisis Persamaan Respon Dosis Thermoluminescent Dosimeter (TLD) Pada Spektrum Sinar-X Menggunakan Metode Monte Carlo

UPAYA MINIMISASI LIMBAH RADIOAKTIF DENGAN CARA PENGAMBILAN KEMBALI RADIONUKLIDA

PENINGKATAN SISTEM PROTEKSI RADIASI DAN KESELAMATAN KAWASAN NUKLIR SERPONG TAHUN 2009

STANDAR NASIONAL INDONESIA (SNI) BIDANG NUKLIR

ANALISIS UPTAKE TIROID MENGGUNAKAN TEKNIK ROI (REGION OF INTEREST) PADA PASIEN HIPERTIROID

KAJIAN KESELAMATAN PADA PROSES PRODUKSI ELEMEN BAKAR NUKLIR UNTUK REAKTOR RISET

PENGARUH SUDUT GANTRI TERHADAP KONSTANSI DOSIS SERAP DI AIR PESAWAT TELETERAPI Co-60 XINHUA MILIK RUMAH SAKIT dr. SARJITO YOGYAKARTA

UJI KESESUAIAN PESAWAT CT-SCAN MEREK PHILIPS BRILIANCE 6 DENGAN PERATURAN KEPALA BAPETEN NOMOR 9 TAHUN 2011

PEMBELAJARAN - 2 PERTEMUAN KE 4 3 x pertemuan DIKLAT FUNGSIONAL PENERA 2011

EVALUASI UNJUK KERJA NERACA MIKRO SEBAGAI IMPLEMENTASI SISTEM MUTU LABORATORIUM PUSAT TEKNOLOGI NUKLIR BAHAN DAN RADIOMETRI

PENGUKURAN AKTIVITAS ISOTOP 152 Eu DALAM SAMPEL UJI PROFISIENSI MENGGUNAKAN SPEKTROMETER GAMMA

Kajian Soal Tertulis Sertifikasi Personel PPR Tahun 2015 Review of the Written Test Question for RPO Sertification Year of 2015

PEMANTAUAN RADIOAKTIVITAS DEBU DI UDARA DAERAH KERJA PPGN TAHUN 2011

EVALUASI DOSIS RADIASI INTERNAL PEKERJA RADIASI PT-BATAN TEKNOLOGI DENGAN METODE IN-VITRO

PENINGKATAN PELAYANAN KESELAMATAN KERJA DI PUSAT TEKNOLOGI NUKLIR BAHAN DAN RADIOMETRI

KOMPARASI PENGUKURAN LAJU KERMA UDARA PESAWAT OB-85 MENGGUNAKAN ALAT UKUR RADIASI STANDAR SEKUNDER DAN STANDAR TURUNANNYA

EVALUASI MESIN STANDAR TORSI NASIONAL PUSLIT METROLOGI - LIPI PADA RENTANG UKUR 5 N M 50 N M

RIWAYAT REVISI /09/2016 Penerbitan Pertama MT MM /10/2016 Perubahan format IK. MT MM

Verifikasi Keluaran Radiasi Pesawat Linac (Foton Dan Elektron) Serta 60CO Dengan TLD

Analisis Pengaruh Sudut Penyinaran terhadap Dosis Permukaan Fantom Berkas Radiasi Gamma Co-60 pada Pesawat Radioterapi

SIMULASI KURVA EFISIENSI DETEKTOR GERMANIUM UNTUK SINAR GAMMA ENERGI RENDAH DENGAN METODE MONTE CARLO MCNP5

EVALUASI KINERJA DOSE CALIBRATOR CAPINTEC CRC-55tR UNTUK PENGUKURAN AKTIVITAS RADIOISOTOP 175 Yb

ANALISIS UNSUR RADIOAKTIVITAS UDARA BUANG PADA CEROBONG IRM MENGGUNAKAN SPEKTROMETER GAMMA

BAB I PENDAHULUAN Latar Belakang

KAJIAN PENERIMAAN DOSIS RADIASI PADA PEKERJA D.A.LAM PROSES DISMANTLIN(3 SUMBER BEKAS

UJI KESTABILAN PENCACAH RADIASI DOSE CALIBRATOR

R adalah selisih massa bejana dalam keadaan terisi dan dalam keadaan kosong,

PENGEMBANGAN SILABUS PELATIHAN DALAM RANGKA PENINGKATAN KOMPETENSI PETUGAS PROTEKSI RADIASI BIDANG MEDIS

STANDARDISASI 1251 DAN PENENTUAN WAKTU PARO SERTA RESPON KAMAR PENGION

Analisis tingkat kontaminasi permukaan daerah kerja dan laju paparan radiasi pada Instalasi Kedokteran Nuklir

Widyanuklida, Vol. 16 No. 1, November 2017: ISSN

KOMPARASI PRAKIRAAN DOSIS INTERNA SECARA IN-VIVO DAN IN-VITRO. R. Suminar Tedjasari, Ruminta Ginting, Tri Bambang L Pusat Teknologi Limbah Radioaktif

KAJIAN LAJU PAPARAN RADIASI PADA TITIK PENGUKURAN DI REAKTOR KARTINI SEBAGAI DASAR PENENTUAN KONDISI BATAS OPERASI (KBO)

KAnANPERANANSUMBERSTANDAR RADIONUKLIDA DI RUMAH SAKIT

PRA RANCANGAN KONTAINER TEMPAT PENYIMPANAN LIMBAH RADIOAKTIF SUMBER TERBUNGKUS 192 Ir

PENGARUH DIAMETER PHANTOM DAN TEBAL SLICE TERHADAP NILAI CTDI PADA PEMERIKSAAN MENGGUNAKAN CT-SCAN

PADA DOSE CALIBRATOR VICTOREEN

PENINGKATAN KESTABILAN PENGUKURAN FREKUENSI RUBIDIUM DENGAN PEMANFAATAN GLOBAL POSITIONING SYSTEM DISCIPLINED OSCILLATOR

STUDI PENGELOLAAN LIMBAH RADIOAKTIF PADAT PEMBANGKIT LISTRIK TENAGA NUKLIR

SISTEM MANAJEMEN KESELAMATAN RADIASI

Transkripsi:

YOGYAKARTA, 16 NOVEMBER 011 PENENTUAN CALIBRATOR SETTING CAPINTEC CRC-7BT UNTUK SAMARIUM-153 Wijono, Gatot Wurdiyanto Pustek Keselamatan dan Metrologi Radiasi - BATAN, Jl.Lebak Bulus No.49 Jakarta, 1440 johnrida@batan.go.id ABSTRAK Penentuan Calibrator Setting Capintec CRC-7BT untuk Samarium-153. Telah dilakukan penentuan calibrator setting sistem Radionuclide Calibrator Capintec CRC-7BT untuk Samarium-153 (Sm-153) di Pusat Teknologi Keselamatan dan Metrologi Radiasi Badan Tenaga Nuklir Nasional. Penelitian ini dilakukan untuk mengetahui nilai calibrator setting, faktor kalibrasi dan ketidakpastian bentangan Capintec CRC-7BT untuk pengukuran aktivitas Sm-153. Metode pengukuran yang digunakan adalah relatif. Sumber standar radionuklida yang digunakan adalah Sm-153 kode A15301/11 dengan aktivitas 1795,5 Ci pada saat pengukuran. Sumber standar tersebut dalam bentuk cair dengan volume (3,95 ± 0,%) ml pada wadah ampul gelas 5 ml. Proses penelitian diawali dengan pencacahan latar dan sumber standar dengan masingmasing pengambilan data 15 kali. Pencacahan sumber standar dilakukan dalam rentang calibrator setting antara 30 sampai 39 yang terbagi dalam 10 tingkat dengan interval 1. Dengan demikian diperoleh rerata, deviasi standar dan ketidakpastian tipe A dari masing-masing tahap pengukuran. Nilai ketidakpastian tipe B diperoleh dari umur paro Sm-153 sebesar 0,006%, di samping faktor-faktor lain yang terdiri dari kebocoran, resolusi, respon, linieritas, akurasi dan repeatability dari alat ukur radiasi yang secara berturutturut sebesar 0,5%; 0,033%; %, %, 3% dan 1%. Hasil perhitungan menunjukkan ketidakpastian gabungan dan bentangan Capintec CRC-7BT masing-masing sebesar,10% dan 4,508%. Hasil menunjukkan calibrator setting Capintec CRC-7BT untuk pengukuran Sm-153 sebesar 34 ± 1 dengan aktivitas (1799,3 ± 0,04%) Ci. Dari pembagian hasil aktivitas peluruhan dengan hasil pengukuran sumber standar diperoleh nilai Faktor Kalibrasi. Hasil perhitungan Faktor kalibrasi Capintec CRC-7BT untuk Sm- 153 sebesar (0,998 ± 4,508%). Dengan hasil ini diharapkan PTKMR BATAN sebagai acuan laboratorium nasional dalam bidang metrologi radiasi mampu melakukan kalibrasi dan standardisasi untuk berbagai macam alat ukur aktivitas dan jenis radioisotop dengan selamat dan aman dari segi proteksi radiasi, efektif, efisien, cepat dan tingkat akurasi memadai. Kata kunci : calibrator setting, faktor kalibrasi dan radionuclide calibrator ABSTRACT Determination of calibrator setting of Capintec CRC-7BT for Samarium-153 (Sm-153) in the Center for Technology of Radiation Safety and Metrology of the National Nuclear Energy Agency has been carried out. This research was conducted to determine the value of setting calibrator, calibration factor and expended uncertainty of Capintec CRC-7BT for activity measuring of Sm-153. The measurement method used is relative. Radionuclide standard source used is the Sm-153 code A15301/11 with 1795.5 Ci activity at the time of measurement. Standard sources in the form of liquid with a volume (3.95 ± 0.%) ml in 5 ml glass ampoule container. How committed is the background counting and a standard source with each data aquitition 15 times. Standard sources counting performed in a range of calibrator setting around 30 to 39, divided into 10 levels with intervals of 1. Thus obtained the mean, standard deviation and type A uncertainty of each measurement phase. Value of type B uncertainty is obtained from the half-life of Sm-153 at 0.006%, in addition to other factors which consist of leakage, resolution, response, linearity, accuracy and repeatability of radiation measuring instrument are 0.5 %, 0.033%, %, %, 3% and 1% respectively. The calculation results show the combined and expanded uncertainty of CRC-7BT Capintec is,10% and 4,508% respectively. Results showed calibrator setting of Capintec CRC-7BT for Sm-153 measuring is 34 ± 1 with (1799.3 ± 0.04%) Ci activity. From the division of activity decay with the measurement results of the standard source obtained calibration factor value. Calibration factor calculation results of Capintec CRC-7BT for Sm-153 was (0.998 ± 4.508%). With this result is expected the Center for Technology of Radiation Safety and Metrology of the National Nuclear Energy Agency as national laboratories reference in the field of radiation metrology capable of performing calibration and standardization for a variety of measuring devices and types of radioisotope activity with a safe and secure in terms of radiation protection, effective, efficient, speedy and adequate level of accuracy. Keywords : calibrator setting, calibration factor and radionuclide calibrator Sekolah Tinggi Teknologi Nuklir-BATAN 98 Wijono dkk

YOGYAKARTA, 16 NOVEMBER 011 1. PENDAHULUAN Radionuclide Calibrator Capintec CRC-7BT merupakan salah satu jenis alat ukur radiasi standar milik Pusat Teknologi dan Keselamatan Radiasi yang digunakan untuk pengukuran aktivitas radionuklida, salah satunya adalah Samarium-153 (Sm-153). Alat tersebut memiliki 8 jenis tombol radionuklida dan beberapa hasil setting radionuklida yang memiliki Calibrator Setting (CS) yang berbeda-beda sesuai jenisnya. Masing-masing tombol memiliki Faktor Kalibrasi yang belum tentu sama antara yang satu dengan lainnya. Sebagai alat ukur standar Capintec CRC-7BT harus memiliki nilai Faktor Kalibrasi yang tertelusur oleh sumber standar sesuai jenis radionuklida yang akan diukur. Masa berlaku Faktor Kalibrasi/kalibrasi ulang untuk Radionuclide Calibrator ini selama 1 tahun (sesuai Peraturan Kepala Badan Pengawas Tenaga Nuklir No.1 tahun 006 tentang Laboratorium Dosimetri, Kalibrasi Alat Ukur Radiasi dan Keluaran Sumber Radiasi Terapi dan Standardisasi Radionuklida). Faktor kondisi dan kualitas alat dapat menyebabkan Faktor Kalibrasi dari Radionuclide Calibrator pada tombol Sm-153 mengalami perubahan. Apabila perubahannya cukup besar maka akan mempengaruhi akurasi hasil pengukurannya. Pada kondisi ini peluang terjadinya kesalahan pengukuran semakin besar, terutama untuk sampel aktivitas kecil. Oleh karena itu perlu dilakukan penentuan Calibrator Setting (CS) Radionuclide Calibrator Capintec CRC-7BT secara berkala untuk memperbaiki kontinuitas nilai Faktor Kalibrasinya. Untuk menghindari kesalahan perhitungan dalam penentuan Faktor Kalibrasi diperlukan koreksi peluruhan terhadap aktivitas sampel karena umur paro Sm-153 cukup pendek, yaitu (1,9855 ± 0,00005) hari. [1] Tujuan dari penentuan Calibrator Setting Radionuclide Calibrator Capintec CRC-7BT pada Tombol Sm-153 adalah untuk memperoleh nilai Calibrator Setting, Faktor Kalibrasi dan ketidakpastian bentangannya. Dengan demikian diharapkan dapat meningkatkan kualitas hasil kalibrasi terhadap alat-alat ukur radiasi yang lain yang biasanya dimiliki oleh Rumah Sakit, sehingga pihak konsumen (pasien) lebih mendapatkan perlindungan keselamatan dari radiasi sesuai peraturan ketenaganukliran yang berlaku. [] detektor (u res ), linieritas (u lin ), akurasi (u akr ) dan repeatability (u rpb ). [3] Untuk tipe A, deviasi standar pengukuran capintec ( ) adalah [4] N i1 (A i A N(N 1) ) (1) sehingga ketidakpastian Capintec (u ) dalam prosen (%) adalah u A 100. (N 1) () Dari nilai-nilai ketidakpastian tipe A dan B diperoleh nilai ketidakpastian standar gabungan (u c ) u c u u u u u u. (3) leak pr resl Apabila k adalah faktor cakupan untuk nilai derajat kebebasan efektif (v eff ) dengan tingkat kepercayaan 95% dan menggunakan nilai kritis k- students, maka nilai ketidakpastian bentangan (U) dapat ditentukan. [3] res i (4) Nilai ketidakpastian bentangan (U) diperoleh dari hasil perkalian nilai faktor cakupan (k) dengan nilai ketidakpastian gabungan (u c ). U = ku c. (5) Tabel 1 dibahas di mana pada teks? Tanda desimal titik diganti koma. Di hal. 4 juga ada Tabel 1. Tabel 1. Hubungan derajat kebebasan terhadap faktor cakupan.. TINJAUAN PUSTAKA Perhitungan nilai ketidakpastian (U) terdiri dari tipe A dan B. Tipe A meliputi nilai-nilai ketidakpastian data pengukuran alat ukur standar Capintec CRC-7BT (u ) dan tipe B meliputi nilainilai ketidakpastian kebocoran detektor (u leak ), umur paro teori (u pr ), resolusi bacaan (u resl ), respon Wijono dkk 99 Sekolah Tinggi Teknologi Nuklir-BATAN

3. TATA KERJA Sistem peralatan yang digunakan dalam penelitian ini meliputi: Radionuclide Calibrator Capintec CRC-7BT S/N 7174, Monitor 4 S/N 6140, Stabiliser Model SKU-3000-VA S/N 4068337374 dan satu set alat proteksi radiasi (sarung tangan karet, TLD, pinset, kontainer Pb dan tissue. Sumber radionuklida standar yang digunakan adalah Sm-153 Kode A15301/11 dengan aktivitas 1795,5 Ci pada saat pengukuran. Sumber standar tersebut dalam bentuk cair dengan volume (3,95 ± 0,%) ml pada wadah ampul gelas 5 ml. Skema penentuan Calibrator Setting Radionuclide Calibrator Capintec CRC-7BT secara lebih lengkap ditunjukkan dalam Gambar 1. Metode pengukuran yang digunakan adalah relatif. Proses ini diawali dengan pencacahan latar dan sumber standar dengan masing-masing pengambilan data sebanyak 15 kali. Pencacahan sumber standar dilakukan dalam rentang calibrator setting antara 30 sampai 39 yang terbagi dalam 10 tingkat dengan interval 1. Dengan demikian diperoleh rerata, deviasi standar dan ketidakpastian tipe A dari masing-masing tahap pengukuran. Nilai ketidakpastian tipe B diperoleh dari umur paro Sm- 153 sebesar 0,006%, di samping faktor-faktor lain SEMINAR NASIONAL YOGYAKARTA, 16 NOVEMBER 011 yang terdiri dari kebocoran, resolusi, respon, linieritas, akurasi dan repeatability dari alat ukur radiasi yang secara berturut-turut sebesar 0,5%; 0,033%; %, %, 3% dan 1%. Dari perhitungan ketidakpastian tipe A dan B dapat ditentukan ketidakpastian gabungan dan bentangan Capintec CRC-7BT Dari pembagian hasil aktivitas peluruhan dengan hasil pengukuran sumber standar diperoleh nilai Faktor Kalibrasi dengan toleransi sama dengan nilai ketidakpastian bentangan. Karakteristik perubahan fungsi calibrator setting terhadap aktivitas (dalam satuan mci) ditentukan dengan interval 1 dari 11 tingkat pengukuran. Dari proses ini diperoleh hasil grafik Faktor Radionuklida (RF) versus Aktivitas (mci) dari Radionuclide Calibrator tersebut. Nilai faktor kalibrasi yang ideal dengan nilai 1,000 terletak pada pertemuan garis karakteristik terhadap nilai aktivitas sumber standar pada saat pengukuran. Apabila hal ini tidak dapat dicapai maka nilai calibrator setting yang diambil adalah titik yang terdekat dengan perpotongan garis grafik seperti tersebut di atas. Setelah ditemukan nilai calibrator setting yang baru secara tepat dan benar maka dilakukan penguncian pengaturan posisi menu pada Radionuclide Calibrator CRC-7BT. Gambar 1. Skema penentuan Calibrator Setting Radionuclide Calibrator Capintec CRC-7BT Sekolah Tinggi Teknologi Nuklir-BATAN 300 Wijono dkk

YOGYAKARTA, 16 NOVEMBER 011 4. HASIL DAN PEMBAHASAN Dari proses pengaturan sistem calibrator setting Capintec CRC-7BT dan berdasarkan data hasil pengukuran aktivitas secara terkoreksi sebagai alat ukur standar diperoleh grafik karakteristik hubungan antara calibrator setting (Cal-Set) versus Aktivitas (Ci) seperti yang ditunjukkan dalam Gambar. Dari gambar tersebut dapat diketahui bahwa semakin kecil pengaturan angka Cal-Set pada Capintec CRC-7BT akan menyebabkan aktivitas hasil pengukuran yang semakin besar (berbanding terbalik). Karakteristik garis grafiknya berbentuk linier walaupun masih terlihat ada beberapa belokan kecil. Garis tersebut berpotongan dengan garis horisontal yang merupakan fungsi aktivitas sumber standar pada saat pengukuran. Titik perpotongan garis tersebut merupakan titik kesetaraan antara hasil pengukuran aktivitas sumber standar menggunakan Capintec CRC-7BT terhadap aktivitas yang sebenarnya sebagai sumber standar yang nilainya telah dikoreksi berdasarkan rumus peluruhan pada saat pengukuran. Apabila pada titik perpotongan ditarik garis vertikal ke bawah dan tepat mengenai angka calibrator setting, maka akan diperoleh Faktor Kalibrasi sebesar 1,00. Namun hal ini tidak terjadi pada Radionuclide Calibrator CRC-7BT karena angka calibrator setting yang dicapai merupakan pembulatan, yaitu 34. Posisi ini terlihat jelas pada grafik Calibrator Setting (CS) versus Selisih Aktivitas Sm-153 (Ci) untuk masingmasing tahap pengukuran seperti yang ditunjukkan dalam Gambar 3. Gambar. Grafik Calibrator Setting (CS) versus Aktivitas Sm-153 (Ci) dari hasil pengukuran Capintec CRC-7BT terhadap sumber standar Gambar 3 Grafik Calibrator Setting (CS) versus Selisih Aktivitas Sm-153 (Ci) untuk masing-masing tahap pengukuran Wijono dkk 301 Sekolah Tinggi Teknologi Nuklir-BATAN

YOGYAKARTA, 16 NOVEMBER 011 Tabel. Ketidakpastian pengukuran Sm-153 menggunakan Capintec CRC-7BT Dari akuisisi data ukur diperoleh rerata, deviasi standar dan ketidakpastian tipe A. Sedangkan nilai ketidakpastian tipe B diperoleh dari umur paro Sm- 153 sebesar 0,006%, di samping faktor-faktor lain yang terdiri dari kebocoran, resolusi, respon, linieritas, akurasi dan repeatability dari alat ukur radiasi yang secara berturut-turut sebesar 0,5%; 0,033%; %, %, 3% dan 1%. Hasil perhitungan menunjukkan ketidakpastian gabungan dan bentangan Capintec CRC-7BT masing-masing sebesar,10% dan 4,508%. Distribusi ketidakpastian Tipe A dan B secara lebih lengkap ditunjukkan dalam Tabel. Hasil menunjukkan calibrator setting Capintec CRC-7BT untuk pengukuran Sm-153 sebesar 34 ± 1 dengan aktivitas (1799,3 ± 0,04%) Ci. Dari pembagian hasil aktivitas peluruhan dengan hasil pengukuran sumber standar diperoleh nilai Faktor Kalibrasi. Hasil perhitungan Faktor kalibrasi Capintec CRC-7BT untuk Sm-153 sebesar (0,998 ± 4,508%). Dengan meninjau kembali Dokumen: IK- LMR-STD-10 maka terlihat jelas bahwa Faktor Kalibrasi Radionuclide Calibrator CRC-7BT memiliki nilai mendekati 1,000. Dengan hasil ini diharapkan PTKMR BATAN sebagai acuan laboratorium nasional dalam bidang metrologi radiasi mampu melakukan kalibrasi dan standardisasi untuk berbagai macam alat ukur aktivitas dan jenis radioisotop dengan selamat dan aman dari segi proteksi radiasi, efektif, efisien, cepat dan tingkat akurasi memadai. 5. KESIMPULAN Semula Radionuclide Calibrator Capintec CRC- 7BT untuk pengukuran aktivitas Sm-153 belum memiliki ketertelusuran terhadap sumber standarnya. Dari hasil ini diperoleh grafik karakteristik dari Calibrator Setting (CS) versus aktivitas Sm-153 dalam satuan Ci. Sedangkan angka Calibrator Setting dan Faktor Kalibrasi Capintec CRC-7BT untuk pengukuran Sm-153 masing-masing sebesar (34 ± 1) dan (0,998 ± 4,508%). Dengan hasil ini diharapkan PTKMR BATAN sebagai acuan laboratorium nasional dalam bidang metrologi radiasi mampu melakukan kalibrasi dan standardisasi untuk berbagai macam alat ukur aktivitas dan jenis radioisotop dengan selamat dan aman dari segi proteksi radiasi, efektif, efisien, cepat dan tingkat akurasi memadai khususnya untuk Sm- 153. 6. UCAPAN TERIMA KASIH 1. Bapak Kepala Pusat Teknologi Keselamatan dan Metrologi Radiasi Badan Tenaga Nuklir Nasional, Jakarta. Bapak Kepala Bidang Metrologi Radiasi PTKMR BATAN 3. Para staf Sub Bidang Standardisasi yang telah banyak membantu dalam pelaksanaan penelitian. 7. DAFTAR PUSTAKA [1] TdeR, Table de Radionuclides Atomic and Nuclear Data, Recommended data/table BNM- LNHB/CEA Table de Radionuclides CEA, http://www. nucleide.org/ddepwg/ddepdata.htm, 004 [] Peraturan Pemerintah, Keselamatan Radiasi Pengion dan Keamanan Sumber Radioaktif, PP No. 33, Jakarta, 007 [3] International Atomic Energy Agency, Measurement Uncertainty, A Practical Guide for Secondary Standards Dosimetry Laboratories, Tecdoc-1585, Vienna, 008 [4] Instruksi Kerja Unit Standarisasi, No. Dokumen: IK-LMR-STD-10, P3KRBiN-Batan, 003. Sekolah Tinggi Teknologi Nuklir-BATAN 30 Wijono dkk